2.16 WEIGH SYSTEM CONTROL PANEL GUIDE (OPTIONAL)
N
N
O
O
T
T
E
E
When the bed is switched on, the scale display will show "STRYKER / SCALE V x.x" where x.x is the
version of the control software.
The Weigh system is intended to assist in the monitoring of the patient's weight variation. Under no
circumstances should its reading be used as sole reference for medical treatment.
A:
key:
•
Press the key once to activate
the display. The Scale mode will
be displayed.
•
Press when instructed to press
ENTER.
B/C:
MENU Up/Down keys:
•
Press either key repeatedly until
desired mode is reached.
D:
Scale Display.
Scale Menu Modes
Order of appearance of the modes when the MENU Down key is pressed repeatedly:
Mode
Display First Line
Scale
WEIGHT ANGLE
Change Equipment
CHANGE EQUIP.
Change Patient Weight
CHNG PTNT WEIGHT
Select Weight Unit (lb/kg)
UNITS
Scale Zero
SCALE ZERO
Operating (Zeroing) the Scale Before Putting a New Patient on the Bed
•
Prepare the bed for a patient stay (linens, pillows, etc.) and press the key, the display will
read:
WEIGHT ANGLE
XXX.X lb +/- XX.X°
•
Press the MENU Up key to access the Scale Zero mode (as written on the membrane). The
display will read:
SCALE ZERO / PRESS ENTER
•
Press and hold the ENTER key. The display will read:
HOLD TO ZERO WT. (keep pressing the ENTER key), followed by:
RELEASE TO ZERO (release the ENTER key), followed by:
DO NOT TOUCH BED (be sure nobody touches the bed while this message appears).
•
The scale system will return to the Scale mode and will display zero for the weight and angle
values. The bed is now ready for the patient.
N
N
O
O
T
T
E
E
The scale display turns off automatically after one minute of idle time, but the system itself will
remains active in the background. Pressing the key will reactivate the display in the Scale mode.
On beds equipped with the Weigh and Bed Exit systems, zeroing one system will also zero the
other.
Do not zero the bed while a patient is on the bed. An inaccurate patient weight reading will result. If
this should occur, use the Change Patient Weight mode to enter the correct patient weight or
remove the patient from the bed and zero the bed again (see above “Operating (Zeroing) the Scale
Before Putting a New Patient on the Bed”).
/
ENTER
W
W
A
A
R
R
N
N
I
I
N
N
G
G
Figure 2.16
D
B
C
A