Stryker BERTEC GOBED PLUS FL20E Operation Manual Download Page 19

 

Chapter 2 

Operation Guide 

 

 

 

 

 

 

 

 

 

 

 

 

 

19

Adding or Removing Equipment when a Patient is on the Bed 

N

N

O

O

T

T

E

E

 

 

Take note of the patient's weight before proceeding with the Change Equipment mode. 
The patient's weight should remain the same at the end of the procedure. If it is not the case, access 
the Change Patient Weight mode using the MENU Up or Down keys and enter manually the 
previously noted patient weight. 

• 

Press the      key to activate the scale display.  

• 

Access the Change Equipment mode using the MENU Down key, the display will read: 

CHANGE EQUIP / PRESS ENTER 

• 

Press and hold the ENTER key. The display will read: 

RELEASE TO START (release the ENTER key), followed by: 

 

DO NOT TOUCH BED (be sure no one touches the bed while this message appears). 

 

• 

The ADD/REMOVE EQUIP / PRESS ENTER message will then appear, signalling that 

equipment can now be added or removed from the bed. 

When done, press and hold the ENTER key. 

• 

Display will read RELEASE TO FIN. (release the ENTER key), followed by: 

• 

DO NOT TOUCH BED (ensure the patient stands still as much as possible and no one 

touches the bed while this message is displayed). 

• 

The scale will return to the Scale mode displaying the patient unchanged weight. 

Changing Manually the Numerical Value of Displayed Weight 

In certain circumstances, it may be necessary to change manually the displayed patient's weight. To 
do so: 

• 

Press the      key to activate the scale display. 

• 

Access the Change Patient Weight mode using the MENU Down key. The display will read: 

CHNG PTNT WEIGHT / PRESS ENTER 

• 

Press and hold the ENTER key. The display will read: 

REL. TO CHNG WT. (release the ENTER key), followed by: 

DO NOT TOUCH BED (be sure no one touches the bed while this message appears). 

• 

The display will then read 

pq

 XXX.X lb / press ENTER. Adjust the weight using the MENU 

Up/Down keys. 

• 

Once done press the ENTER key. 

• 

The scale will return to the Scale mode and will display the adjusted weight. 

The Change Patient Weight mode should not be used systematically for new patients. The scale 
should be zeroed each time a new patient enters onto the bed (see page 18, "Operating (Zeroing) 
the..).

 

Changing the Weight Measure Unit 

• 

Press the      key to activate the scale display. Access the Select Weight Unit (lb/kg) mode 
using the MENU Down key. The display will read: 

UNITS / PRESS ENTER 

• 

Press the ENTER key. The display will read: 

• 

Select the weight unit using the MENU Up or Down key and press the ENTER key. The scale 
will return to the Scale mode with the weight in the unit chosen.

 

pq

  lb AND 

PRESS ENTER 

pq

  kg AND 

PRESS ENTER 

OR 

Summary of Contents for BERTEC GOBED PLUS FL20E

Page 1: ...Canada E mail Canada only service bertec strykercorp com 1 800 327 0770 United States Manufactured by Stryker Bertec Medical Inc 72 0439 R1 3 September 2003 F15 43 E A Printed in Canada OPERATIONS MANUAL ELECTRIC ACUTE CARE BED Model FL20E USA 2501 ...

Page 2: ......

Page 3: ...rm Operation 13 2 6 Patient Restraint Strap Locations 13 2 7 Emergency CPR Release Optional 14 2 8 Night Light Usage Optional 14 2 9 Nurse Call Usage Optional 14 2 10 Auxiliary Power Outlet Usage Option Available w 120V Bed Model 14 2 11 Positioning Siderails 15 2 12 Head and Foot Board Operation 15 2 13 Head Siderail Function Guide 16 2 14 Foot Board Control Panel Guide 17 2 15 Function Lockout U...

Page 4: ...Height 14 5 to 28 9 36 8 to 73 4 cm Fowler Angle w CPR Mechanism Fowler Angle w o CPR Mechanism 0 to 60 0 to 65 Knee Gatch Angle w o Auto Contour Knee Gatch Angle w Auto Contour 0 to 32 0 to 24 Trendelenburg Reverse Trendelenburg 14 to 14 Environmental Conditions Transport and storage Ambient Temperature Relative humidity Atmospheric Pressure Operating Ambient Temperature Relative humidity Atmosph...

Page 5: ...mattress thickness should not exceed 6 25 15 9 cm This bed is equipped with a hospital grade plug for protection against shock hazard It must be plugged directly into a properly grounded receptacle Grounding reliability can be achieved only when a hospital grade receptacle is used Serious injury can result if caution is not used when operating the bed Operate the bed only when all people and equip...

Page 6: ... armed the Bed Exit system may cause a reduction in the sensitivity of the Bed Exit system resulting in erroneous readings of the patient s movements in the bed The Bed Exit system is not designed to be used with patients weighing less than 50 lb 23 kg When large fluid spills occur in the area of the circuit board cables and motors immediately unplug the bed power cord from the wall outlet Remove ...

Page 7: ... reserves the right to charge shipping and restocking fees on returned items DAMAGED MERCHANDISE Claims for damaged merchandise must be made with the carrier within fifteen 15 days of receipt of merchandise DO NOT ACCEPT DAMAGED SHIPMENTS UNLESS SUCH DAMAGE IS NOTED ON THE DELIVERY RECEIPT AT THE TIME OF RECEIPT Upon prompt notification Stryker Bertec will file a freight claim with the appropriate...

Page 8: ...ed _____ Depress the green AXIAL STEER side of the pedal at both sides of the bed fully down to put the bed in the steer mode see page 12 Ensure the 5th steer wheel is properly engaged Toggle the pedal to the neutral position and ensure the 5th wheel disengages _____ Ensure the siderails raise lock in the up position and lower smoothly see page 15 _____ Run through each control of the foot board c...

Page 9: ... mild detergent Wipe the bed clean and dry thoroughly to avoid build up of cleaning solution N NO OT TE E Do not use cleaning solutions containing a degreaser near the siderail pivots see figure 2 2B on page 11 of the Maintenance Manual to avoid deteriorating the grease used to ensure a smooth movement of the siderails Mattress Care Inspect the mattress after each use Discontinue use if any cracks...

Page 10: ...rol motor resets itself automatically as soon as the fowler is completely lowered ____ Verify the Fowler Knee Gatch and Hi lo movements to ensure that the motion interrupt switch integrated to each of the four actuators is operating properly ____ Auxiliary outlet option available only with 120V electric system beds working properly see page 14 ____ Optional night light working properly ____ Foot p...

Page 11: ...ITION CONTOUR POSITION Figure 1 9 POWER CORD CONNECTOR AND ON OFF SWITCH MODULE Figure 1 10 QE71 0299 OPTIONAL EMERGENCY CRANK STORAGE LOCATION OPTIONAL 37 PIN NURSE CALL CABLE CONNECTOR OPTIONAL GEN III PENDANT CONTROL CABLE CONNECTOR FOLEY BAG HOOK OPTIONAL NURSE CALL SPEAKERPHONE SIDERAIL INNER CONTROL PANEL FOOT BOARD CONTROL PANEL BRAKE STEER PEDAL CPR RELEASE HANDLE SIDERAIL RELEASE LEVER SI...

Page 12: ...the red TOTAL BRAKE side of the brake steer pedal at either side of the bed To disengage the wheel brakes toggle the pedal to the neutral position 2 3 MOVING THE BED The bed is equipped with a fifth steer wheel activated by a pedal see page 11 Bed Illustration located at midpoint of the bed on both sides The fifth steer wheel helps guiding the bed along a straight line and helps the bed pivot arou...

Page 13: ...imple Knee Gatch raise is required To do so perform the following steps 1 Before raising the Knee Gatch reach under the foot section lift the prop rod and hold it while raising the Knee Gatch through the foot board control panel 2 As the Knee Gatch raises the rod will come to pass the support brackets Release it then and continue to raise the Knee Gatch The foot section will then simply follow the...

Page 14: ...icate through speakerphones with the nurse station by simply pressing a nurse call red button integrated to both head end siderail inner control panels see page 16 The communication between the patient and the nurse station is established the moment a member of the nursing staff responds to the call signal Included with this option are two Ø1 4 phono jacks located on both sides of the head end cas...

Page 15: ...utward and swing it upward towards the head end of the bed until it latches in the up position To raise a foot end siderail the same procedure is required as for the head end siderail however the siderail swings to the foot end of the bed To lower a siderail lift it up slightly push in the release lever located under the mattress support see page 11 Bed Illustration and hold siderail as it rotates...

Page 16: ... Siderail Patient s Right Rail Inside Siderail Patient s Left Rail A Press to raise Fowler C Press to raise Knee Gatch B Press to lower Fowler D Press to lower Knee Gatch E Press to activate the Nurse Call function optional F Nurse call speakerphones this panel is optional equipment and will be present on the inner side of both head siderails if the bed is equipped with the Nurse Call option see p...

Page 17: ...ain the Trendelenburg position head down foot up E2 Press to obtain the reverse Trendelenburg position head up foot down N NO OT TE E To replace the sleep surface to horizontal position after a Trendelenburg positioning simply use the bed height controls to either raise or lower the sleep surface to its limits Then position the bed to the desired height F Contour positioning control The Contour po...

Page 18: ...splay will read WEIGHT ANGLE XXX X lb XX X Press the MENU Up key to access the Scale Zero mode as written on the membrane The display will read SCALE ZERO PRESS ENTER Press and hold the ENTER key The display will read HOLD TO ZERO WT keep pressing the ENTER key followed by RELEASE TO ZERO release the ENTER key followed by DO NOT TOUCH BED be sure nobody touches the bed while this message appears T...

Page 19: ... the patient unchanged weight Changing Manually the Numerical Value of Displayed Weight In certain circumstances it may be necessary to change manually the displayed patient s weight To do so Press the key to activate the scale display Access the Change Patient Weight mode using the MENU Down key The display will read CHNG PTNT WEIGHT PRESS ENTER Press and hold the ENTER key The display will read ...

Page 20: ...may cause a reduction in the sensitivity of the Bed Exit system resulting in erroneous readings of the patient s movements in the bed The Bed Exit system is not designed to be used with patients weighing less than 50 lb 23 kg N NO OT TE E When the bed is equipped with a scale the scale must be properly zeroed for the Bed Exit to function properly see page 18 Operating Zeroing the Scale Before Putt...

Page 21: ... kg Zone Settings The first zone left LED is the traditional Bed Exit zone The patient can move in the bed freely but cannot fully exit the bed or the alarm will sound The second zone middle LED is more restrictive It allows the patient to sit up and roll over but any attempt to exit the bed will cause the alarm to sound The third zone right LED is the most restrictive zone Small movements like ra...

Page 22: ... with hook Product Design Three function pendant control Neeco Tron Inc Upright oxygen bottle holder Maximum load 75 lb 34 kg Defibrillator tray Maximum load 40 lb 18 kg Sheet cradle Overhead trapeze system Maximum load 150 lb 68 kg 1 Ø removable anodized aluminium IV pole 1 Ø fixed anodized aluminium IV pole 1 2 Ø removable anodized aluminium IV pole Emergency crank Padded siderail covers set of ...

Reviews: