2.7 EMERGENCY CPR RELEASE (OPTIONAL)
The CPR release is for emergency use only. When activating the CPR release, all people and
equipment must be removed from the area below and around the head, thigh and foot sections
of the bed or serious personal injury or equipment damage could occur.
When quick access to the patient is needed, and the Fowler is raised, squeeze one of the
two release handles located under the upper right and left sides of the head section (see
page 11, “Bed Illustration”). The Fowler and the knee section (if also raised) will
automatically flatten. The Fowler control motor will reset itself automatically as soon as the
Fowler will be completely lowered.
The handle can be released at any time to stop the Fowler from lowering. Note that the
resetting of the control motor does not occur when the Fowler is not lowered completely.
N
N
O
O
T
T
E
E
During the short period of time the Fowler control motor is resetting, the bed functions are
not available.
2.8 NIGHT LIGHT USAGE (OPTIONAL)
The bed may be equipped with an optional photoelectric night light to illuminate the floor
area around the bed. The night light turns on as the room lights dim.
2.9 NURSE CALL USAGE (OPTIONAL)
The bed may be equipped with an optional nurse call function allowing the patient to
communicate through speakerphones with the nurse station by simply pressing a nurse call red
button integrated to both head end siderail inner control panels (see page 16).
The communication between the patient and the nurse station is established the moment a
member of the nursing staff responds to the call signal.
Included with this option are two Ø1/4" phono jacks located on both sides of the head end
casing. The phono jacks enable the use of a nurse call cord which can be placed within reach of
a patient who is not in bed.
N
N
O
O
T
T
E
E
The nurse call system will automatically send a call signal to the nurse station if the
communication between the bed and the nurse station is interrupted following a power failure,
the switching off of the bed or the disconnection of the nurse call communication cable.
2.10 AUXILIARY POWER OUTLET USAGE (OPTION AVAILABLE W/120V~ BED MODEL)
Use only hospital grade electric equipments with the auxiliary power outlet. The use of
normal electric equipments may bring the current leakage to a level unacceptable for
hospital equipment.
The North American version (120V~ electrical system) of the may be equipped with
an optional 120V~ auxiliary power outlet integrating a 5A breaker. Located on the left side at the
foot end of the bed (see page 11, “Bed Illustration”), this feature provides nursing staff with a
convenient power source for peripheral equipments.
N
N
O
O
T
T
E
E
Use only equipment with a current consumption of 5 amp or less.