![Stryker 7600-000-050 Maintenance Manual Download Page 66](http://html.mh-extra.com/html/stryker/7600-000-050/7600-000-050_maintenance-manual_1385854066.webp)
Conducted
RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
6 Vrms in ISM bands
150 kHz to 80 MHz
3 V/m
80 MHz to 2.7 GHz
3 Vrms
6 Vrms in ISM bands
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to
any part of A
AR
RG
GA
AIIO
OS
S 2
25
50
0,
including cables, than the
recommended separation
distance calculated from the
equation appropriate for the
frequency of the transmitter.
Recommended separation
distance
D=(1.2) (√
P
)
D=(1.2) (√
P
)
80 MHz to 800 MHz
D=(2.3) (√
P
)
800 MHz to 2.7 GHz
Where
P
is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation
distance in meters (m).
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site survey,
a
should be less than the
compliance level in each
frequency range.
b
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
N
No
otte
e 1
1:: At 80 MHz and 800 MHz, the higher frequency range applies.
N
No
otte
e 2
2:: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
N
No
otte
e 3
3:: The ISM (industrial, scientific, and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz;
13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
N
No
otte
e 4
4:: Evaluated for immunity to proximity fields from RF wireless communication equipment per IEC 60601-1-2:2014
Table 9.
EN
62
SK-7600 REV 06
Summary of Contents for 7600-000-050
Page 2: ......
Page 16: ...S Se er riia all n nu um mb be er r llo oc ca at tiio on n A EN 12 SK 7600 REV 06 ...
Page 21: ...A B C D E F SK 7600 REV 06 17 EN ...
Page 68: ...H HM M 1 17 7 3 31 14 4 A A A HM 17 314 1 EN 64 SK 7600 REV 06 ...
Page 69: ...H HM M 1 17 7 1 16 6 A A HM 17 16 1 SK 7600 REV 06 65 EN ...
Page 70: ...H HM M 1 17 7 1 15 56 6 A A A HM 17 156 1 EN 66 SK 7600 REV 06 ...
Page 71: ...H HM M 1 17 7 3 31 10 0 H HM M 1 17 7 3 31 11 1 A HM 17 310 HM 17 311 1 SK 7600 REV 06 67 EN ...
Page 72: ...H HM M 1 17 7 3 31 13 3 A A HM 17 313 1 EN 68 SK 7600 REV 06 ...
Page 73: ...H HM M 1 17 7 3 31 12 2 A A HM 17 312 1 SK 7600 REV 06 69 EN ...
Page 74: ...H HM M 1 17 7 3 30 04 4 A B A HM 17 304 1 B HM 17 304 1 EN 70 SK 7600 REV 06 ...
Page 75: ...H HM M 1 17 7 3 30 03 3 A A A HM 17 303 1 SK 7600 REV 06 71 EN ...
Page 76: ...H HM M 1 17 7 3 30 05 5 A A HM 17 305 1 EN 72 SK 7600 REV 06 ...
Page 77: ......
Page 78: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA SK 7600 REV 06 WCR AA 11 2021 08 ...