Control box beeps while actuators are
raised and actuators will not move in
the downward direction
Control box actuator position
reference has been lost.
1. Unplug the power cord and turn off
the battery.
2. Plug in the power cord and turn on
the battery.
3. Run each upper actuator to its
highest position.
Failure Mode :
Bed does not operate due to actuator
failure or power request error. When a
failure mode occurs and a key is
activated the system will beep quickly,
and if LEDs are used for locking
functions on Nurse Control Unit, these
will blink quickly, except when
powering down on battery.
Failure indications:
1) All Nurse Control Unit LEDs are
blinking.
2) The CB6 buzzer beeps quickly if the
handset is activated.
1. Actuator cables are unplugged.
2. Electrical short in a control cable
(cable of control units or actuators).
3. If a function is activated but the
actuator configuration doesn’t
support the function.
4. Switch has failed in an actuator.
5. Control pendant (nurse or patient)
or siderail control panel has failed.
6. Control box has failed.
1. Reset the failure mode.
2.
2.1. Press and hold the
Trendelenburg button (A) on
the nurse control panel and
the Autocontour up (B) on the
patient control panel at the
same time for five seconds.
N
No
otte
e -- The Trendelenburg
and Autocontour up buttons
are found on the opposite
sides of the same siderail
control panel.
2.2. Press and hold the Backrest
up (C) and Backrest down
(D) buttons at the same time
on the patient control
pendant (option), or
Autocontour up (E) and
Trendelenburg (F) buttons at
the same time on the nurse
control pendant (option) for
five seconds.
3. After five seconds, you will hear a
buzzer.
4. Keep pressing buttons until the
buzzer stops.
5. LEDs will stop blinking.
6. Press the CPR button in to zero all
the actuators.
7. Failure mode does not result in a
position lost. Make sure that the
system is in a safe position after
resetting the failure mode.
8. Continue to troubleshoot through
possible causes until corrected.
EN
16
SK-7600 REV 06
Summary of Contents for 7600-000-050
Page 2: ......
Page 16: ...S Se er riia all n nu um mb be er r llo oc ca at tiio on n A EN 12 SK 7600 REV 06 ...
Page 21: ...A B C D E F SK 7600 REV 06 17 EN ...
Page 68: ...H HM M 1 17 7 3 31 14 4 A A A HM 17 314 1 EN 64 SK 7600 REV 06 ...
Page 69: ...H HM M 1 17 7 1 16 6 A A HM 17 16 1 SK 7600 REV 06 65 EN ...
Page 70: ...H HM M 1 17 7 1 15 56 6 A A A HM 17 156 1 EN 66 SK 7600 REV 06 ...
Page 71: ...H HM M 1 17 7 3 31 10 0 H HM M 1 17 7 3 31 11 1 A HM 17 310 HM 17 311 1 SK 7600 REV 06 67 EN ...
Page 72: ...H HM M 1 17 7 3 31 13 3 A A HM 17 313 1 EN 68 SK 7600 REV 06 ...
Page 73: ...H HM M 1 17 7 3 31 12 2 A A HM 17 312 1 SK 7600 REV 06 69 EN ...
Page 74: ...H HM M 1 17 7 3 30 04 4 A B A HM 17 304 1 B HM 17 304 1 EN 70 SK 7600 REV 06 ...
Page 75: ...H HM M 1 17 7 3 30 03 3 A A A HM 17 303 1 SK 7600 REV 06 71 EN ...
Page 76: ...H HM M 1 17 7 3 30 05 5 A A HM 17 305 1 EN 72 SK 7600 REV 06 ...
Page 77: ......
Page 78: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA SK 7600 REV 06 WCR AA 11 2021 08 ...