13. Remove the hydraulics assembly and properly discard.
N
No
otte
e -- Do not dispose of as unsorted municipal waste. Refer to your local distributor for return or collection systems
available in your country.
14. Reverse steps to reinstall.
15. Verify proper operation before you return the product to service.
H
Hyyd
drra
au
ulliicc ccyylliin
nd
de
err rro
od
d e
en
nd
d rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
1/4'' hex wrench
•
1/2'' combination wrench
•
Diagonal pliers
P
Prro
occe
ed
du
urre
e::
1. Pull the trolley assembly out of the patient compartment until it locks into the loading position.
2. Remove the trolley covers. See
3. Press the main power button to turn the product off.
4. Using a 1/4" hex wrench and a 1/2" combination wrench, remove the end cap cylinder pin (A) and nut (B) (Figure 25).
EN
44
6390-309-002 Rev AB.0
Summary of Contents for 6390
Page 2: ......
Page 6: ......
Page 85: ...6390 309 002 Rev AB 0 79 EN ...
Page 91: ...6390 309 002 Rev AB 0 85 EN ...
Page 95: ...Torque item B to 68 92 in lb Torque item K to 20 28 ft lb 6390 309 002 Rev AB 0 89 EN ...
Page 96: ...Torque item DE to 480 600 in lb EN 90 6390 309 002 Rev AB 0 ...
Page 97: ...Torque item C to 45 max in oz 6390 309 002 Rev AB 0 91 EN ...
Page 98: ...EN 92 6390 309 002 Rev AB 0 ...
Page 99: ...6390 309 002 Rev AB 0 93 EN ...
Page 100: ...EN 94 6390 309 002 Rev AB 0 ...
Page 101: ...6390 309 002 Rev AB 0 95 EN ...
Page 102: ...EN 96 6390 309 002 Rev AB 0 ...
Page 103: ...6390 309 002 Rev AB 0 97 EN ...
Page 137: ......