5. Reverse steps to reinstall.
6. Verify proper operation before you return the product to service.
M
Mo
otto
orr rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
9/32'' socket
•
1/4'' drive ratchet
P
Prro
occe
ed
du
urre
e::
C
CA
AU
UT
TIIO
ON
N -- Damage may occur if motor armature or stator are bumped around when you replace the motor.
1. Pull the trolley assembly out of the patient compartment until it locks into the loading position.
N
No
otte
e -- This makes working on the product easier, but is not required.
2. Remove the trolley covers. See
3. Remove the motor cable. See
Motor cable removal and replacement
4. Using a 9/32" socket and 1/4" drive ratchet, remove the two bolts (A) that secure the motor (B) to the hydraulic manifold
assembly (C) (Figure 43).
N
No
otte
e -- Do not overtighten the bolts during reinstallation.
6390-309-002 Rev AB.0
57
EN
Summary of Contents for 6390
Page 2: ......
Page 6: ......
Page 85: ...6390 309 002 Rev AB 0 79 EN ...
Page 91: ...6390 309 002 Rev AB 0 85 EN ...
Page 95: ...Torque item B to 68 92 in lb Torque item K to 20 28 ft lb 6390 309 002 Rev AB 0 89 EN ...
Page 96: ...Torque item DE to 480 600 in lb EN 90 6390 309 002 Rev AB 0 ...
Page 97: ...Torque item C to 45 max in oz 6390 309 002 Rev AB 0 91 EN ...
Page 98: ...EN 92 6390 309 002 Rev AB 0 ...
Page 99: ...6390 309 002 Rev AB 0 93 EN ...
Page 100: ...EN 94 6390 309 002 Rev AB 0 ...
Page 101: ...6390 309 002 Rev AB 0 95 EN ...
Page 102: ...EN 96 6390 309 002 Rev AB 0 ...
Page 103: ...6390 309 002 Rev AB 0 97 EN ...
Page 137: ......