1. Remove the patient right side cover. Make sure that the linkages of the manual release that attaches to the side controls
are connected and actuating the manual release valve on the hydraulic pump.
a. If the manual release valve is actuating, then the problem is a faulty velocity fuse. The velocity fuse (6390-001-381)
is located in the cap end cylinder port under the elbow fitting. Remove and replace the velocity fuse.
b. If replacing the velocity fuse does not solve the problem, replace the pressure compensated flow control (6390-001-
151).
c. Replace the hydraulics assembly (6390-001-039).
T
Trro
olllle
eyy m
ma
an
nu
ua
all rre
elle
ea
asse
e b
bu
utttto
on
n llo
ow
we
errss tth
he
e lliiffttiin
ng
g a
arrm
mss,, b
bu
utt n
no
ott ssm
mo
oo
otth
hllyy
1. Remove and replace the velocity fuse (6390-001-381).
N
No
otte
e -- The velocity fuse is located in the cap end cylinder port under the elbow fitting.
2. If this does not fix the issue, replace the pressure compensated flow control (6390-001-151).
3. Replace the hydraulics assembly (6390-001-039).
T
Trro
olllle
eyy e
errrro
orr L
LE
ED
D iin
nd
diicca
atte
ess a
an
n e
errrro
orr ((sso
olliid
d a
am
mb
be
err))
Contact Stryker Technical Support.
T
Trro
olllle
eyy cco
on
nttrro
oll p
pa
an
ne
ell d
do
oe
ess n
no
ott m
mo
ovve
e tth
he
e lliiffttiin
ng
g a
arrm
mss
1. Make sure that the power is on (solid green LED on the trolley control panel).
2. Make sure that the trolley is in the loading position and the latches are engaged.
N
No
otte
e -- The trolley control panel will only work if the trolley is in the loading position and the latches are connected. When
the lifting arms are supporting weight, they only lower to a certain angle. These are safety features.
3. Check the head end indicators (solid green LED) to make sure that the product is in the loading position and that both
latches are engaged.
a. If there is no head end indicator LED activity, then the product is not in the loading position.
b. If the head end indicator LEDs are flashing amber, one or both of the latches are not closed. Check the status of the
latches.
c. If the problem persists, check the switches and connectivity back to the control board.
4. Check connectivity from the trolley control panel back to the control board.
5. Replace the trolley control panel board (6390-001-450).
T
Trro
olllle
eyy sstto
op
pss p
pa
arrtt w
wa
ayy w
wh
hiille
e rro
olllliin
ng
g tto
o tth
he
e ttrra
an
nssp
po
orrtt p
po
ossiittiio
on
n
1. Remove the transfer from the anchor.
a. Check for and remove any debris from the anchor and the channels on the inside of the transfer.
b. Examine the trolley to transfer lock release ramp (6390-001-144) to make sure that it is intact, tightly fastened, and
not excessively worn.
6390-309-002 Rev AB.0
25
EN
Summary of Contents for 6390
Page 2: ......
Page 6: ......
Page 85: ...6390 309 002 Rev AB 0 79 EN ...
Page 91: ...6390 309 002 Rev AB 0 85 EN ...
Page 95: ...Torque item B to 68 92 in lb Torque item K to 20 28 ft lb 6390 309 002 Rev AB 0 89 EN ...
Page 96: ...Torque item DE to 480 600 in lb EN 90 6390 309 002 Rev AB 0 ...
Page 97: ...Torque item C to 45 max in oz 6390 309 002 Rev AB 0 91 EN ...