4. Remove the solenoid valve.
5. Disconnect the connector for the solenoid valve.
N
No
otte
e -- Inspect the replacement solenoid valve to make sure that the O-ring is in place. If the O-ring is not in place,
transfer the O-ring from the removed solenoid valve to the supplied solenoid valve.
6. Discard the solenoid valve.
7. Reverse steps to reinstall.
N
No
otte
e
•
During reinstall, align the V
Ve
ellccrro
o® patches.
•
After installation, remove the zipper pull tool or equivalent from the zipper.
•
Cover the zipper with the support surface cover watershed.
8. Verify proper operation before you return the product to service.
P
Po
ow
we
err b
bo
oa
arrd
d rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
Zipper pull tool or equivalent
•
T10 Torx driver
•
ESD system
P
Prro
occe
ed
du
urre
e::
1. See
2. See
Protecting against electrostatic discharge (ESD)
3. Using a T10 Torx driver, remove the four screws (C) that secure the power board (BB) to the foot box (Figure 12). Save
the screws.
EN
32
2973-009-002 Rev AB.0
Summary of Contents for 297300000000
Page 2: ......
Page 13: ...Product serial number Completed by Date 2973 009 002 Rev AB 0 9 EN ...
Page 52: ...EN 48 2973 009 002 Rev AB 0 ...
Page 53: ...2973 009 002 Rev AB 0 49 EN ...
Page 54: ...EN 50 2973 009 002 Rev AB 0 ...
Page 55: ...2973 009 002 Rev AB 0 51 EN ...
Page 59: ...2973 009 002 Rev AB 0 55 EN ...
Page 60: ...EN 56 2973 009 002 Rev AB 0 ...
Page 61: ...2973 009 002 Rev AB 0 57 EN ...
Page 62: ...EN 58 2973 009 002 Rev AB 0 ...
Page 63: ...2973 009 002 Rev AB 0 59 EN ...
Page 64: ...EN 60 2973 009 002 Rev AB 0 ...
Page 65: ...2973 009 002 Rev AB 0 61 EN ...