0207-050-702 Rev-L
EN
www.stryker.com
3
Features
J
I
E
G
F
A
B
E
E
H
C
D
K
J
L
K
G
M
A
Applied Part(s)
– The tip assembly and handpiece body are
the applied parts as defined by the manufacturer according to
the standards listed under
Product Safety Certification
in the
Specifications
section.
B
Latch Mechanism –
The mechanism is used to release the tip
assembly from the handpiece.
C
Trigger –
An activated trigger provides irrigation to the surgical/wound
site.
D
Trigger Lock –
The lock is used to hold the trigger at either the high
irrigation flow or low irrigation flow setting.
E
Irrigation Tubing/Power Cord Combination –
The irrigation tubing
provides a fluid pathway from the irrigation bag to the handpiece tip
assembly. The power cord conducts electrical energy from the battery
power pack to the handpiece.
F
Suction Clamp –
The clamp is used to open and close the suction
tubing pathway to control suction at the handpiece tip assembly (not
included with all handpiece configurations).
G
Suction Tubing –
The suction tubing provides a suction pathway
from the handpiece tip assembly to the vacuum canister connected
to the facility vacuum source (not included with all handpiece
configurations).
H
Suction Tube Port –
The port allows connection of the suction tubing
to the handpiece (not shown).
I
Battery Power Pack (removable batteries) –
The battery pack
provides power to the handpiece and allows easy removal of internal
batteries. (REF 0207-580-000, REF 0207-583-000, REF 0207-585-
000, REF 0207-586-000, REF 0207-589-000 and REF 0207-590-000
only)
Features (continued)
J
Irrigation Tube Spike –
The spike is used to pierce and connect the
handpiece irrigation tubing to the irrigation bag.
K
Irrigation Tube Spike Cap –
The protective spike cap is installed over
the sharp irrigation tube spike to maintain sterility and prevent damage
or injury.
L
Battery Power Pack (sealed) –
The battery pack provides power to
the handpiece.
M
Suction Tube Adapter –
The adapter is used for a large diameter
port on a vacuum fluid collection canister and may be removed if
necessary.
Definitions
The symbols located on the equipment and/or labeling are defined in this section
or in the
Symbol Definition Chart
. See the
Symbol Definition Chart
supplied with
the equipment.
SYMBOL
DEFINITION
Consult Instructions for Use
General warning sign
Battery lid removal tabs
(REF 0207-580-000, REF 0207-583-000, REF 0207-
585-000, REF 0207-586-000, REF 0207-589-000 and
REF 0207-590-000 only)
Trigger Lock
Type BF Applied Part
Sterilized using irradiation
Direct Current (DC)
Do not re-use
Do not use if package is damaged
Instructions
To Assemble the Handpiece
WARNINGS:
• ALWAYS use aseptic technique when handling the sterile handpiece and
irrigation tubing.
• DO NOT reuse the handpiece for surgical applications or applications
requiring a sterile field.
• ALWAYS use aseptic technique when removing the irrigation tube spike cap.
Failure to comply may cause the irrigation tube spike to become contaminated
before insertion into the irrigation bag.
NOTES:
• Make sure a standard IV pole is equipped with an irrigation bag.
• Make sure the top of the fluid contained in the irrigation bag is less than 36
inches above the handpiece at all times after connecting the handpiece.
0000004063, Rev. L Effective Date: -