TegraPol-11/ -15
TegraForce-1
Instruction Manual
22
To remove TegraForce from TegraPol:
Switch off TegraPol with the switch at the back of the machine.
Disconnect TegraForce from the compressed air supply.
Remove the TegraDoser-1 or the dosing nozzle from
TegraDoser-5 from the TegraForce.
Disconnect the TegraForce from the TegraPol by removing the
plug.
Release the handle and let the TegraForce swing in the upper
position.
Remove the preparation disc.
Press the TegraForce down again until the head is locked in
position.
Remove the two caps covering the fastening screws on the left-
hand side of the TegraPol cabinet by lifting them off with a knife.
Loosen the two fastening screws while holding the TegraForce in
position.
Lift the TegraForce straight up and place it on a safe and stable
support.
Removing TegraForce-1 from
TegraPol
WARNING!
TegraForce is relatively heavy. When loosening the two fastening screws it
might slide downwards. Keep your hands off the column and away from
between specimen mover disc and turntable.