To ensure optimal cooling, TegraPol can be fitted with a Struers
Recirculation Cooling Unit.
Take the inlet hose delivered with TegraPol and cut off the
coupling nut on the straight end.
Slide the hose clamp onto the hose and insert the quick coupling
delivered with the pump for the recirculation cooling unit. Tighten
the hose clamp.
Connect the quick coupling on the inlet hose directly to the cooling
unit’s pump outlet (
A
).
If necessary, use one of the hoses delivered with the pump to
extend the inlet hose.
Mount the other end to the water inlet tube.
Mount the outlet hose onto the water outlet tube. Use a hose
clamp to secure the hose.
If there is only limited space behind the TegraPol, use the pipe
bend and attach it directly to the outlet tube on the back of the
TegraPol. Then attach the drain hose to the other end of this
bend.
Lead the other end into the mounting hole in the bracket on top of
the static filter unit (
B
). Arrange the tube so that it slopes
downward towards the drain throughout its length. Shorten the
hose, if necessary.
Connect the communication cable to the Cooli Control Box and
then connect to the socket
1
at the rear of the TegraPol.
Connecting a Recirculation
Cooling Unit
Please note:
The Recirculation Cooling Unit cannot be used with TegraPol units running
methods that include the dosing of oxide polishing suspensions.
Note
Before connecting the cooling unit to the TegraPol,
follow the instructions in the Struers Cooling Units Instruction Manual
to prepare it for use.
Connecting the Water Inlet
Connecting the Water Outlet
Connecting the Communication
Cable
IMPORTANT
Before connecting, check that the mains voltage corresponds to the voltage
stated on the type plate on the side of the machine.
A
B