7.22 MUX OUT channel select register (0x110)
#define SIS3820_MUX_OUT_CHANNEL_SELECT 0x110 /* read/write; D32 */
MUX functionality (available from firmware 01 0A on) allows to route one of the 32 scaler
input signals to an external instrument like a visual rate meter e.g.
Register Bit assignments:
Bit Function
31
no function, read as 0
... ...
5
no function, read as 0
4
bit 4 of OUT_MUX channel
3
2
1 ...
0
bit 0 of OUT_MUX channel
Notes:
1.) The power up default is0x0 (i.e. scaler channel 0 is routed to the output as soon as output
mode 3 is activated)
2.) MUX_OUT mode requires output mode 3
Example:
Channel 5 (channels counted from 0 to 31) is passed to output 7 with bit 0 and 2 of the MUX
OUT channel select register set and output mode 3 enabled