I. Visual inspection
OK
Not
ok
Description
of Defect
What?
How?
Patient lifting
pole, adapter
sleeves, grab
handle with strap
No damage or deformation
Manufacturer's recommendation: Replace grab handle after 5
years
Bed frame: Mat-
tress base, chas-
sis
No damage, deformation, no split welded
seams
Castors
No damage
Mattress base
No damage or deformation
Safety sides
No damage, cracks or deformation
Connecting ele-
ments (screws,
bolts, nuts, safe-
ty caps)
Secure fixing, missing parts
Wearing parts,
joints
No damage or severe wear
Plastic parts, var-
ious
Free of sharp-edged damage/breakage
II. Electrical Measurement according to DIN EN 62353 (VDE 0751-1)
Limit val-
ue
Measured
value
Earth conductor resistance measuring
point: PE pin on the bed, head end,
(recommended measuring current for
beds cleaned in automatic washing
systems ≥ 2 A)
0.3 Ω
____ Ω
Device leakage current type: B, direct
(place bed with conductive castors in a
way that it is insulated) or difference.
0.5 mA
____ mA
Appendix
151
Instruction Manual
Summary of Contents for Seta pro
Page 1: ...Seta pro 2020 11 12 Version 00 288675 Instruction Manual Model variant junior...
Page 9: ......
Page 101: ...4 Reinsert the handrail on the headboard Use routine Instruction Manual 92...
Page 166: ...Appendix 157 Instruction Manual...
Page 167: ...Appendix Instruction Manual 158...
Page 168: ......