Ports for signal input/signal output parts
Phenomenon
EMC basic standard
Immunity test level
Professional healthcare
facilities
Electrostatic discharge (ESD)
IEC 61000-4-2
+/- 8 kV; contact
+/- 2 kV, +/- 4 kV ¸ +/- 8 kV, +/-
15 kV air
Electrical fast transients/bursts
IEC 61000-4-4
+/- 1 kV; 100 kHz repetition fre-
quency
Conducted interference in-
duced by high-frequency fields
IEC 61000-4-6
3 V; 0.15 MHz to 80 MHz;
6V in ISM frequency bands be-
tween 0.15 MHz and 80MHz
80% AM at 1 kHz
Table zz: Test specifications for the immunity of sheathings to high-frequency wire-
less communication equipment
Test Fre-
quency
MHz
Frequen-
cy band
Radio
service
Modula-
tion
Max.
power W
Distance
m
Immunity
test level
v/m
385
380 to 390
TETRA 400
Pulse mod-
ulation 18
Hz
1.8
0.3
27
450
430 to 470
GMRS 460
FRS 460
+/- 5% lift,
1kHz sine
wave
2
0.3
28
710
704 to 787
LTE band
13, 17
Pulse mod-
ulation 217
Hz
0.2
0.3
28
745
780
810
800 to 960
GSM
800/900
TETRA 800
iDEN820,
CDMA 850,
LTE band 5
Pulse mod-
ulation 18
Hz
1
0.3
28
870
930
Appendix