1-4
10083002
Operator Manual
Safety Precautions
Appropriate components of this lighting system have been tested and found in compliance with IEC
60601-1-2:2007, Medical Electrical Equipment – Part 1: General Requirements for Safety;
Electromagnetic Compatibility (EMC). There is, however, a potential for electromagnetic or other
interference between this equipment and other devices. Should you experience interference, relocate this
device or minimize the use of the affected equipment while this device is in operation.
Do not bump lightheads into walls or other equipment. Always use handles when positioning lighthead
during surgical procedures, or when cleaning or servicing the lighting system.
Portable and mobile RF communications equipment may temporarily affect the operation of the
HarmonyAIR Surgical Lighting System equipment.
Use of ACCESSORIES, transducers and cables other than those specified, with the exception of
transducers and cables sold by the manufacturer of this device as replacement parts for internal
components, may result in increased EMISSIONS or decreased IMMUNITY of the HarmonyAIR Surgical
Lighting System.
The HarmonyAIR Surgical Lighting System should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the equipment or system should be observed to verify
normal operation in the configuration in which it is used.
Accessories or replacement parts not listed in the Operator or Maintenance Manuals should not be used
as it may affect EMC or result in equipment damage.
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put
into service according to the EMC information provided in this manual.
To avoid inadequate balancing, load added to single monitor yoke must not exceed 33 lb (15 kg).
To avoid inadequate balancing, load added to dual monitor yoke must not exceed 26 lb (12 kg) per
monitor.
CAUTION – POSSIBLE EQUIPMENT DAMAGE HAZARD: (CONTINUED)
!
!
!
!
!
!
!
!
!
Summary of Contents for Harmony LED Wall Control Unit
Page 2: ......
Page 8: ...vi 10083002 Operator Manual Table of Contents Healthcare Capital Equipment ...
Page 10: ...viii 10083002 Operator Manual Table of Contents Healthcare Capital Equipment ...
Page 18: ...2 4 10083002 Operator Manual Symbol Definitions Healthcare Capital Equipment ...
Page 66: ...8 2 10083002 Operator Manual Maintenance Healthcare Capital Equipment ...
Page 68: ...9 2 10083002 Operator Manual Replacement Parts Healthcare Capital Equipment ...
Page 70: ...10 2 10083002 Operator Manual Waste Management Guidelines Healthcare Capital Equipment ...