![Steris Harmony LED Wall Control Unit Operator'S Manual Download Page 12](http://html.mh-extra.com/html/steris/harmony-led-wall-control-unit/harmony-led-wall-control-unit_operators-manual_1374339012.webp)
1-2
10083002
Operator Manual
Safety Precautions
Monitor spring arms must be locked (using built-in safety pin) prior to installing and/or removing the
monitor and/or monitor yoke.
WARNING – POSSIBLE PATIENT INJURY HAZARD:
Failure to engage the lighthandle cover completely may result in cover falling from lighthead during the
procedure.
WARNING – BIOHAZARD:
Sterile disposable covers are intended for single use only.
Universal disposal precautions must be observed when disposing of any single use disposable item.
WARNING – PERSONAL INJURY HAZARD:
When fiber optic video accessories are installed to the HarmonyAIR Surgical Lighting System, do not
operate system unless optical fiber cables are fully connected.
To maintain Hazard Levels (1M or less) and safety measures established by STERIS, end user must
assure that any laser device integrated into the system and supplied by sources other than STERIS does
not exceed the optical radiation power (per wavelength) of corresponding STERIS transmitters listed in
Table 1-1. It is the responsibility of the end user to ensure any device not supplied by STERIS
corresponds to the values listed in Table 1-1.
Do not attempt to clean lighthead unless power is turned off and the lighthead has cooled sufficiently.
Do not attempt to adjust suspension system. Refer servicing to qualified service personnel.
Avoid looking directly at into illuminated lighthead. Looking directly into high-intensity light may result in
eye injuries.
Do not attempt to replace LED modules on HarmonyAIR M Series lights. Refer servicing to qualified
service personnel.
WARNING – IMPACT HAZARD: (CONTINUED)
!
!
!
!
!
!
Table 1-1. System Laser Components
Type of Laser
Source
Max. Output
/ Channel
Channels
Wavelength
Notes
(dBm)
(mW)
(nm)
SDI Transmitter
+7dBm
5.01
1
1310
Sources of radiation - self-
contained,
Class 1
Laser
Products
S-Video Transmitter
0dBm
1.0
1
1310
VGA Transmitter
+3dBm
2.0
3
1310
DVI Transmitter
-3.6dBm
0.44
4
850
!
!
!
!
Summary of Contents for Harmony LED Wall Control Unit
Page 2: ......
Page 8: ...vi 10083002 Operator Manual Table of Contents Healthcare Capital Equipment ...
Page 10: ...viii 10083002 Operator Manual Table of Contents Healthcare Capital Equipment ...
Page 18: ...2 4 10083002 Operator Manual Symbol Definitions Healthcare Capital Equipment ...
Page 66: ...8 2 10083002 Operator Manual Maintenance Healthcare Capital Equipment ...
Page 68: ...9 2 10083002 Operator Manual Replacement Parts Healthcare Capital Equipment ...
Page 70: ...10 2 10083002 Operator Manual Waste Management Guidelines Healthcare Capital Equipment ...