STATUS SCIENTIFIC CONTROLS
Installation, Commissioning & Routine Gas Testing
FGD10A-M Gas Detector
TD18/032
Issue:
10
Change Note:
1869
Page 63
9.1.27.
Set Flammable Gas Default LEL and Correction Factor
NOTE: This option is for use by SSCL only and is password protected.
This option restores the factory-set default LEL and propane correction
factors for all gases on the Flammable list.
Selecting ENTER for this option will produce following warning:
Default LEL&CF.
Edit
WARNING [ENTER] WILL
RESTORE DEFAULT LEL&
CORRECTION FACTORS
FOR ALL FLAM. GASES
[MENU] FOR NO CHANGE
As stated on the screen, selecting MENU will exit without any changes.
Selecting ENTER key will reset all propane correction factors and LEL
settings to their factory-set default values. These will be displayed on a
series of screens that update consecutively every few seconds.
Screen 1: Screen 2:
Default LEL&CF.
Data updated
GAS CF Lel
Methane ---- 5.0
Propane ---- 2.0
Butane 0.97 1.6
Chloromethane 4.97 8.1
Cyclopentane 1.62 1.1
Default LEL&CF.
Data updated
GAS CF Lel
Ethane 1.01 2.0
Ethanol 1.95 2.1
Ethylene 3.43 2.7
Ethylene Ox 0.85 3.6
Hexane 0.80 1.1
Screen 3:
Default LEL&CF.
Data updated
GAS CF Lel
Pentane 0.89 2.1
Propylene 1.69 2.4
Toluene 1.16 1.2
Xylene 1.32 1.0
User 1.00 1.0
Note: This will also reset the User defined option back to 1.0 CF and LEL.