7.2 Maintaining
Multi-Channel Pipettes
O-rings for the pipette cone are removed and replaced using the
optional auxiliary tool (see section 8 Ordering Information).
Disassembling the lower
housing of the pipette
Fig. 9
Fig. 10
•
Fully unscrew the lower housing
16
by
turning clockwise (Fig. 9).
•
To remove the tip ejector
5
squeeze with
finger and thumb on either side of the tip
ejector, just below the ejector fixing
6
. This
will allow you to slightly lift the ejector fixing
6
which will release the tip ejector
5
.
•
Pull both clips
17
down and rotate the lower
housing through 180° (Fig. 10).
The figures shown below refer to the 100µl model. They may differ
slightly for other pipette models.
Changing the O rings
(100µl and 300µl models only)
18