16
ErgoOne
®
User Manual
Replacing seals for
models up to 1000µl
•
Remove the spacer ring
11
(only
100µl, 200µl and 1000µl models)
(Fig. 5).
•
Remove the sealing ring
10
from the
tip holder using the auxiliary tool
(Fig. 6) and replace it (see section 8
Ordering Information for replace-
ment sealing ring).
•
Reassemble the pipette in the
reverse order.
•
Remove the O-ring
7a
, spring
8
and piston
9
(Fig. 5).
Note:
The piston is subject to spring tension.
•
Degrease the piston
9
using an alcoholic
disin fectant and re-lubricate the piston
with original STARLAB piston grease (see
section 8 Ordering Information).
A damaged piston must be replaced.
•
Reassemble the pipette in the reverse
order.
Fig. 5
Fig. 6