1.2 Front-Panel Operation
1 – 13
1.2.5.4 Store and Recall
Press the [Store] key, followed by a number
i
and the [Enter] key to
store the current configuration in memory location
i
. The parameter
i
may range from 1 to 9. Location 0 is reserved for current instrument
settings and is overwritten after each key press.
The following settings are saved:
1. Whether the shutter is in
sleep
mode
2. Shutter polarity
3. Trigger mode
4. All trigger delay settings and step sizes
5. Alarm Mute
/
Audible selection
6. Display on
/
o
ff
state
Press the [Shift][Recall] key, followed by a number
i
and the [Enter]
key to recall and restore configuration
i
. The parameter
i
may range
from 0 to 9.
1.2.5.5 Alternate shutter modes
SR470 Series Laser Shutters may have alternate operation modes
available, for example alternate speed modes. The possible modes
are numbered 0 through 3, and their function is hardware dependent;
please see the shutter head operation manual for details.
To configure the shutter to operate in an alternate mode, press
[Shift][Align].
The numeric display will display (for example)
“
” to indicate that alternative mode 0 is currently selected.
The arrow keys or numeric keypad can then be used to change the
value.
Important note: Alternate mode settings are stored in the volatile
memory of the shutter head, not in the SR470; the mode setting is
lost whenever the SR470 or the shutter head is unplugged, powered
down, or otherwise reset.
1.2.5.6 Display Blanking
Press [Shift][Disp. O
ff
] to temporarily blank the front panel indica-
tors. (This function is not actually in the “Utility” section, but is
instead in the numeric keypad area.)
When blanking is turned on, all front panel LEDs will go dark except
for the
Disp O
ff
enunciator located to the left of the numeric display.
To turn the display back on, press any key.
SR470 Shutter Controller