5-2
Lead Tests
Pacing, Mode, and V. Pacing Parameter Autoprogramming
The Pacing, Mode, and V. Pacing parameters are linked, so that a change to one parameter results in
automatic changes to the other two. For example, enabling Pacing in the atrial channel (when
ventricular pacing is off) changes the Mode setting from ODO to AAI. Similarly, changing the Mode
from AAI to DDD will turn on Pacing in the ventricular channel (or in a CRT device, the ventricular
channels designated by the V. Pacing parameter). In CRT devices. changing the V. Pacing setting from
RV Only to Simultaneous enables pacing in the LV channel (if the Mode parameter is set to DDD or
VVI).
The following table above shows how the Merlin PSA is programmed for all combinations of the
Pacing, V. Pacing, and Mode parameters.
Accessed From: Lead Tests button
CAUTION
Leaving the atrial connector disconnected when the Merlin PSA is in
DDD or VDD modes may result in the unit sensing external noise
and increasing the ventricular pacing rate to the Max Track Rate.
A Pacing
RV Pacing
LV Pacing
Mode
Setting
V. Pacing
Setting
Off
Off
Off
ODO
n/a
Off
Off
On
VVI
LV Only
Off
On
Off
VVI
RV Only
Off
On
On
VVI
Biventricular
On
Off
Off
AAI
n/a
On
Off
On
DDD
LV Only
On
On
Off
DDD
RV Only
On
On
On
DDD
Biventricular
Table 5-1. Changes to Mode and V. Pacing settings when pacing is enabled (On) or disabled
(Off) in a particular channel