7-4
Parameters
V. Sensing
The V. Sensing parameter determines the channel in which ventricular sensing occurs when the
selected device implant type is CRT-D or CRT-P.
When the
parameter is changed, the V. Sensing parameter is autoprogrammed to match the
chamber selected for V. Pacing.
See also:
•
(radio buttons on Lead Tests window)
•
Parameter Settings and Availability
Ventricular Support Rate (VOO)
The Ventricular Support Rate (VOO) parameter determines the stimulation rate applied to the ventricle
during atrial Burst Pacing.
See also:
•
Parameter Settings and Availability
•
CAUTION
For CRT devices, if only one ventricular channel is connected to the
Merlin PSA, ensure that the Ventricular Sensing button on the Lead
Tests window corresponds to the connected channel. Failure to do
so will result in asynchronous pacing.