If the
'Automatic wash'
option has been checked, it can be started as soon as the injection cycle has been
completed on all loops. Depending on the method, there may be several washing sequences.
6.2
Temperatures
The temperature of the transfer lines is self-regulated at 250 °C and therefore cannot be changed.
The valves oven temperature can be changed in the
Method\Temperatures
menu.
The following window appears:
The temperature can be set between 150 and 250 °C. Click on "Apply" to send values to the storage
interface, then click on "Ok" and save in the IST16 method.
Note that when the temperature from 150 to 250 °C is applied, the storage interface will start its own
heating procedure and will automatically trigger the switch valves during the temperature raise. This
operation is done to preserve the system from mechanical problems with the rotary valves.
6.3
Loops selection
For some experiments, it is not necessary to use all the 16 storage loops. In this case, you can disable the
loops which will not be used from the
Interface\Select loops
menu.
It is strongly recommended to work only with consecutive loops to keep the performance in terms of
reproducibility of results.