Step 3:
Legend :
Inactive Active
In this configuration of the IST16, the leak may be:
-
On the selected loop
→
change the loop and check if the leak has disappeared, replace it if
necessary.
-
On the rotor of the storage valve
→
clean the rotor (paragraph 10.3); replace it
if necessary.
-
On the connection between the isolation valve and the storage valve
→
check for leaks and
replace tube and ferrules if necessary.
-
On the atmospheric pressure release valve, inspect the plastic vent located on the rear panel
→
if
you detect helium at the tube outlet in this configuration, it means that the atmospheric pressure
valve is leaking. Change the atmospheric pressure valve.
Contact SRA Instruments or your TGA support company for the next steps.