In the IST16 oven, located between the rear panel of the IST16 and the injection valve
→
check for
leaks and replace tube and ferrules if necessary.
-
On the rotor of the injection valve
→
clean the rotor (see paragraph 10.3); replace it if necessary.
-
If the GC holds the pressure in this position, go to step 2.
Step 2:
Legend :
Inactive Active
In this configuration of the IST16, the leak may be:
-
Between the injection valve connection and the isolation valve connection
→
check for leaks and
replace tube and ferrules if necessary.
-
On the rotor of the isolation valve
→
clean the rotor (paragraph 10.3); replace it
if necessary.
-
If the GC holds the pressure in this position, go the step 3.