16.
Appendix III: Installation checkout sample
16.1
Sample and conditions
•
Sample:
Calcium carbonate (CaCO
3
)
•
Measuring cell:
TGA/DSC 3+ with SDTA sensor,
IST16 storage interface,
Agilent 7890A GC / 5975C MSD.
•
Pre-treatment:
None
•
Sample preparation: See respective figures
16.2
Instrument parameters
16.2.1
TGA
•
Reactive/Purge gas: 30 mL/min, N
2
grade 5.0
•
Protective gas: 20 mL/min, N
2
grade 5.0
•
Method: 500 °C to 1000 °C at 20 °C/min.
•
Pan: Alumina 70 µL
•
Sample mass: around 10 mg
16.2.2
IST16
Transfer lines and valves temperature: 250 °C
16.2.3
GC
•
Inlet port temperature: 280 °C
•
Split: 3:1
•
Septum purge flow: 1 mL/min
•
Column specifications: HP-5 60 m x 0.32 mm x 0.25 µm
•
Column flow: 0.8 mL/min.
•
Oven temperature program: 50 °C for 4 min, 20 °C/min. to 100 °C, isothermal at 100 °C for 1 min.
16.2.4
MSD
•
Scan mode: m/z from 35 to 55
•
EMV gain: 1