System Care
©
2016-2017 SR Research Ltd.
103
Fingerprints,
smudges,
routine
disinfection.
3
•
Wipe the window and/or enclosure with pre-
packaged 70% isopropyl alcohol swabs. DO NOT
USE saturated cotton balls or >70% strength
alcohol.
•
Let dry thoroughly before allowing participants near
the unit.
Grease, oily
or dried
deposits.
4
•
Mix mild detergent in water, apply to cloth and wring
out so cloth is damp (not wet). Wipe front to back
several times. Rinse cloth and wring out, and wipe
unit several times to remove detergent and dirt.
•
Let dry thoroughly before allowing participant in
area.
Thorough
disinfection
(exposure to
disease
agents)
5
•
Prepare a solution of 10% household bleach (< 0.6%
sodium hypochlorite).
•
Saturate and wring out cloth (must be damp, not
dripping).
•
Wipe front to back on enclosure and left to right on
window.
•
DO NOT LET BLEACH SOLUTION DRY ON UNIT.
Wipe several times with cloth dampened with water.
•
Let dry thoroughly before allowing participant in
area.
Table 3: Approved Cleaning and Disinfecting Methods, Ordered by Risk