GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS
The Body Beautiful is intended for use in the electromagnetic environment specified below. The customer or the user of
the Body Beautiful should assure that it is used in such an environment.
EMISSIONS
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
RF emissions
CISPR 11
Group 1
The Body Beautiful uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The Body Beautiful is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Conform
Class A
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Conform
f
f
GUIDANCE AND MANUFACTURER’S DECLARATION
ELECTROMAGNETIC IMMUNITY
The Body Beautiful is intended for use in the electromagnetic environment specified below. The customer or the user of
the Body Beautiful should assure that it is used in such an environment.
IMMUNITY
TEST
IEC 60601 TEST LEVEL
COMPLIANCE LEVEL
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
Not applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) and neutral
Not applicable
Voltage
dips, short
interruptions
and voltage
variations
on power supply
input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
Not applicable
Mains power quality should be that of a typical
commercial or hospital environment. If a dips or an
interruption of mains power occurs, the current of
the Body Beautiful may be dropped off from normal
level, it may be necessary to use uninterruptible
power supply or a battery
40 % U
T
(60 % dip in U
T
)
for 5 cycles
Not applicable
70 % U
T
(30 % dip in U
T
)
for 25 cycles
Not applicable
<5 % U
T
(>95 % dip in U
T
)
for 5s
Not applicable
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
NOTE - U
T
is the a.c. mains voltage prior to application of the test level