43
Door removal
The following procedure describes how to remove the sliding door. This may be
necessary to resolve door operating issues (e.g. adjust the height of the locking
solenoid, access the safety switch...)
IMPORTANT
: Always disconnect the instrument from the power outlet when
working with panels removed. 240 Volts inside! Risk of electrocution!
1.
Disconnect the instrument from the wall outlet.
2.
Remove the Front panel of the unit (see previous section).
3.
Remove the screws that fasten the door handle.