Date format:
year, month, day
Time format:
hour, minute, second
Alarm threshold:
1-999µSv/h, default value 205µSv/h
Pulse sound:
ON/OFF
Time setting for average radiation measurement: the setup can
change the processing reaction time of the meter against radiation
source with setup range from 8s to 120s. With the increasing of radi
-
ation dose rate, it will automatically and proportionally reduce the av
-
erage time according to the current setting average time. When the
setting time is 8s and the radiation strength is more than SµSv/h, the
fastest response time can reach is 2s. Factory default time is 30s.
Bluetooth data transmission: “ON” and “OFF”.
6-9. Date Setup
1. Press “setup” button once, select “DATE” icon by pressing “UP” or
“DOWN” buttons, and the icon will be in flashing state. Press ENTER
button to confirm your choice, and press ENTER button again for
setup of “year”, “month”, and “day”. For example, when “year” icon
is flashing, press “UP” or “DOWN” button to adjust the value, then
press “enter” to confirm.
2. Press “ENTER” button and set “month”, when “month” icon is
flashing, press “UP” or DOWN” to adjust the value, then confirm the
setup by pressing ENTER.
3. Press ENTER for the 3rd time to set “day”, the method is the
same as above.
4. After finishing setting, press SETUP and return to main menu,
select the next setup or press “ESC” to exit.
6-10. Time Setup
Press “setup” button once, select “TIME” icon through “UP” or
“DOWN”. Same method of DATE setup as above( hour, minute,
second).
12
6-11. Alarm Threshold Value Setup (default value is 205uSv/h)
1. Press “SETUP” button once, then select “ALM” (BL) icon through
“UP” or “DOWN” button. Then, the “ALM” icon will flash, press EN
-
TER button to confirm your option, and press ENTER button once
again to set the single digit, ten digit and hundred digit. The related
digit will flash, and can be adjusted by pressing “UP” or “DOWN” but
-
ton. After completing the setting, press “SETUP” and return to main
menu.
2. When the measurement value is more than your set value, there
will be continuous beep sound.
6-12. Pulse sound ON&OFF Setup
1. Press “setup” button once, and select icon by pressing “UP”
and “DOWN” button. The icon will be flashing, and press ENTER
button to confirm your option. it will display “OFF” and “ON” and in
flashing sate, and then select pulse sound ON or OFF state via “UP”
and “DOWN” button, and press ENTER to confirm your choice. After
the setting, press SETUP to return to main menu, select next menu
or press “ESC” button to exit.
2. When radiation ray has been detected, there will be a “tick” sound.
And the stronger the radiation signal is, the faster the response
frequency will be.
6-13. Setup of average time of radiation measurement (default
value is 30s)
1. Press “setup” button once, select “T” icon by “UP” or “DOWN”
button, and “T” icon will be flashing. Press ENTER to confirm your
option, press ENTER button again to adjust the time, single, tens,
and hundreds that flashing, by “UP” or “DOWN” button.
2. After the setting, press SETUP button to return to the main menu,
select the next menu or press “ESC” button to exit.