Date format:

 year, month, day

Time format:

 hour, minute, second

Alarm threshold:

 1-999µSv/h, default value 205µSv/h

Pulse sound:

 ON/OFF

Time setting for average radiation measurement: the setup can 

change the processing reaction time of the meter against radiation 

source with setup range from 8s to 120s. With the increasing of radi

-

ation dose rate, it will automatically and proportionally reduce the av

-

erage time according to the current setting average time. When the 

setting time is 8s and the radiation strength is more than SµSv/h, the 

fastest response time can reach is 2s. Factory default time is 30s.
Bluetooth data transmission: “ON” and “OFF”.

6-9. Date Setup

1. Press “setup” button once, select “DATE” icon by pressing “UP” or 

“DOWN” buttons, and the icon will be in flashing state. Press ENTER 

button to confirm your choice, and press ENTER button again for 

setup of “year”, “month”, and “day”. For example, when “year” icon 

is flashing, press “UP” or “DOWN” button to adjust the value, then 

press “enter” to confirm.
2. Press “ENTER” button and set “month”, when “month” icon is 

flashing, press “UP” or DOWN” to adjust the value, then confirm the 

setup by pressing ENTER.
3. Press ENTER for the 3rd time to set “day”, the method is the 

same as above.

4. After finishing setting, press SETUP and return to main menu, 

select the next setup or press “ESC” to exit.

6-10. Time Setup

Press “setup” button once, select “TIME” icon through “UP” or 

“DOWN”. Same method of DATE setup as above( hour, minute, 

second).

12

6-11. Alarm Threshold Value Setup (default value is 205uSv/h)

1. Press “SETUP” button once, then select “ALM” (BL) icon through 

“UP” or “DOWN” button. Then, the “ALM” icon will flash, press EN

-

TER button to confirm your option, and press ENTER button once 

again to set the single digit, ten digit and hundred digit. The related 

digit will flash, and can be adjusted by pressing “UP” or “DOWN” but

-

ton. After completing the setting, press “SETUP” and return to main 

menu.

2. When the measurement value is more than your set value, there 

will be continuous beep sound.

6-12. Pulse sound ON&OFF Setup

1. Press “setup” button once, and select     icon by pressing “UP” 

and “DOWN” button. The     icon will be flashing, and press ENTER 

button to confirm your option. it will display “OFF” and “ON” and in 

flashing sate, and then select pulse sound ON or OFF state via “UP” 

and “DOWN” button, and press ENTER to confirm your choice. After 

the setting, press SETUP to return to main menu, select next menu 

or press “ESC” button to exit.
2. When radiation ray has been detected, there will be a “tick” sound. 

And the stronger the radiation signal is, the faster the response 

frequency will be.

6-13. Setup of average time of radiation measurement (default 

value is 30s)

1. Press “setup” button once, select “T” icon by “UP” or “DOWN” 

button, and “T” icon will be flashing. Press ENTER to confirm your 

option, press ENTER button again to adjust the time, single, tens, 

and hundreds that flashing, by “UP” or “DOWN” button. 
2. After the setting, press SETUP button to return to the main menu, 

select the next menu or press “ESC” button to exit.

13

Summary of Contents for 840023

Page 1: ...MULTI FUNCTION DIGITAL RADIATION DETECTOR 840023 ...

Page 2: ...ion data transmission and powerful software for data evaluation It can detect a β y and X ray It also can detect radiation pulse times This meter can be widely used in pharmaceutical factory s laboratory s power stations stone pits emergency rescue stations metal plants oil fields and the oil pipelines environmental protections police depart ments and etc This meter can detect Groundwater radium p...

Page 3: ... confirm button 9 ESC exit button in any setting or mea suring conditions press this button can return to Setup state 10 Power consumption measurement of batter and parameter setting up key 11 SV unit accumulating count key and timing measuring key 12 CPS CPM pulse counting measurement unit switch button 4 4 LCD display symbol description 1 Boot nuclear radiation ray measuring dynamic icon indicat...

Page 4: ...European CE US FCC 15 Warranty One year for the main engine 6 7 Note Sievert mark is Sv unit 1Sv 100 rem rem roentgen equivalent man µSv h unit of ray radiation dose rate 1SV 100Rem 1 rem 1cSv 10mSv 1uR h 10uSv h 1 mR h 10mSv h CPS pulse times per second In the pulse rate model pulse has constantly been measured by counter tube and translated into pulse rate CPM pulse counts per minute ln the puls...

Page 5: ...on The digital nuclear radiation scanner could also do the measure ment recording the accumulated radiation dose at specific time The initial unit is µSv and it will automatically convert to mSv or Sv as the measured dose rate getting higher 6 5 Button for Dose Rate accumulation 1 Press SV button and enter into the regular infinite dose rate accumulation calculation mode with dose rate icon µSv on...

Page 6: ...into timing pulse counting mode and shows 10 the pulse and timing flashing icon on display press this button again to stop pulse counting pressing once again will return to CPS pulse counting function 3 Long press CPS CPM button for two seconds or under pulse measurement mode and will enter into measuring time setup mode and the screen shows pulse and time icon At this time the dis play shows thre...

Page 7: ... method of DATE setup as above hour minute second 12 6 11 Alarm Threshold Value Setup default value is 205uSv h 1 Press SETUP button once then select ALM BL icon through UP or DOWN button Then the ALM icon will flash press EN TER button to confirm your option and press ENTER button once again to set the single digit ten digit and hundred digit The related digit will flash and can be adjusted by pr...

Page 8: ...cally record the measure ment data during the transmission For the connection with PC pis refer to the software operation instruction 7 Battery Voltage Detection function 1 The multi function digital nuclear radiation scanner uses sup per powerful Lithium battery Press UP or BATTERY DETECTION button it will display the inter nal battery voltage normal voltage is 2 7V to 3 7V 2 Press ESC button to ...

Page 9: ...ble This warranty does not cover probes batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will void the warranty To obtain warranty service ship the unit postage prepaid to SPER SCIENTIFIC LTD 8281 E Evans Rd Suite 103 Scottsdale AZ 85260 The defective unit must be accompanied by a description of the pro...

Reviews: