background image

Attention: 

Be careful when rotating the switch to avoid damage to   

the sensor.
Under normal conditions, the meter can detect the current radiation 

value quickly and accurately when  the sensor window of rays aim at 

radiation source.

6-2. Power source and back light switch button.

 

Note:

 The stored data will still be saved after turn it off, but the 

stored data will be lost if it is turned off Inappropriate.

1. Press the switch button for 2 seconds, it will beep shortly, and ac

-

tivate the LCD display; Press again the switch button for 2 seconds 

to shut down the LCD display.

2. After turn it on, and press the switch button, the LCD back light 

will be activated, press again, the back light will be shut off. To save 

energy, the back-light will be automatically shut off 1 minutes after 

be activated.

6-3. Sv/h and Rem/h button for current radiation dose rate

1. Press the button and into standard mode, it displays the current 

measurement value (µSv/h). The measurement results could be 

read in data mode, and bar graph as well. The bar graph on the 

lower corner of display will change as the reading fluctuating and is 

distributed in scales as 0.1,1,10,100. Note: If the reading is too low, 

the bar graph will be show in a line.
2. Sv/h and Rem/h unit conversion
Press the button for conversion, which is based on the formula: 

10µSv/h = 1mRem/h

8

6.4. Sv Dose accumulation

The digital nuclear radiation scanner could also do the measure

-

ment, recording the accumulated radiation dose at specific time. The 

initial unit is µSv, and it will automatically convert to mSv or Sv as the 

measured dose rate getting higher.

6-5. Button for Dose Rate accumulation

1. Press “SV” button, and enter into the regular infinite dose rate 

accumulation calculation mode with dose rate icon ( µSv ) on the 

display. Press the button again, and enter into timing measurement 

mode (default value is 60 minutes). Press the button once again to

stop the timing measurement mode with a beep warning, and it dis

-

plays the current accumulated measurement value; Press the button 

one more time, it will return to regular infinite dose rate accumulation 

calculation mode.

2. 

Under the timing measurement mode, press this button 

for 2 seconds for setting the measuring time, then it displays dose 

icon (µSv) and time unit icon “min”. Now, the display will indicate 3 

numbers as 060 as default value, which means 60 minutes, and the 

maximum is 999 minutes.

The first digit(the hundreds) will be in flashing state for setting

directly.

press         Enter button once to set tens digit ;
press         Enter button the second time to set single digit;
press         ESC button and return to the previous state;
press         Enter button for the 3rd time to confirm the time setup 

and shows the current setting value;
press         Enter button for the 4th time into dose accumulative 

working state, and the icon “TIME” on the top left corner starts flash

-

ing. press       down button or       up button to adjust the display

value.

9

Summary of Contents for 840023

Page 1: ...MULTI FUNCTION DIGITAL RADIATION DETECTOR 840023 ...

Page 2: ...ion data transmission and powerful software for data evaluation It can detect a β y and X ray It also can detect radiation pulse times This meter can be widely used in pharmaceutical factory s laboratory s power stations stone pits emergency rescue stations metal plants oil fields and the oil pipelines environmental protections police depart ments and etc This meter can detect Groundwater radium p...

Page 3: ... confirm button 9 ESC exit button in any setting or mea suring conditions press this button can return to Setup state 10 Power consumption measurement of batter and parameter setting up key 11 SV unit accumulating count key and timing measuring key 12 CPS CPM pulse counting measurement unit switch button 4 4 LCD display symbol description 1 Boot nuclear radiation ray measuring dynamic icon indicat...

Page 4: ...European CE US FCC 15 Warranty One year for the main engine 6 7 Note Sievert mark is Sv unit 1Sv 100 rem rem roentgen equivalent man µSv h unit of ray radiation dose rate 1SV 100Rem 1 rem 1cSv 10mSv 1uR h 10uSv h 1 mR h 10mSv h CPS pulse times per second In the pulse rate model pulse has constantly been measured by counter tube and translated into pulse rate CPM pulse counts per minute ln the puls...

Page 5: ...on The digital nuclear radiation scanner could also do the measure ment recording the accumulated radiation dose at specific time The initial unit is µSv and it will automatically convert to mSv or Sv as the measured dose rate getting higher 6 5 Button for Dose Rate accumulation 1 Press SV button and enter into the regular infinite dose rate accumulation calculation mode with dose rate icon µSv on...

Page 6: ...into timing pulse counting mode and shows 10 the pulse and timing flashing icon on display press this button again to stop pulse counting pressing once again will return to CPS pulse counting function 3 Long press CPS CPM button for two seconds or under pulse measurement mode and will enter into measuring time setup mode and the screen shows pulse and time icon At this time the dis play shows thre...

Page 7: ... method of DATE setup as above hour minute second 12 6 11 Alarm Threshold Value Setup default value is 205uSv h 1 Press SETUP button once then select ALM BL icon through UP or DOWN button Then the ALM icon will flash press EN TER button to confirm your option and press ENTER button once again to set the single digit ten digit and hundred digit The related digit will flash and can be adjusted by pr...

Page 8: ...cally record the measure ment data during the transmission For the connection with PC pis refer to the software operation instruction 7 Battery Voltage Detection function 1 The multi function digital nuclear radiation scanner uses sup per powerful Lithium battery Press UP or BATTERY DETECTION button it will display the inter nal battery voltage normal voltage is 2 7V to 3 7V 2 Press ESC button to ...

Page 9: ...ble This warranty does not cover probes batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will void the warranty To obtain warranty service ship the unit postage prepaid to SPER SCIENTIFIC LTD 8281 E Evans Rd Suite 103 Scottsdale AZ 85260 The defective unit must be accompanied by a description of the pro...

Reviews: