background image

1

Bluetooth Datalogging 

Thermo-Hygrometer

800020

Instruction Manual 

Summary of Contents for 800020

Page 1: ...1 Bluetooth Datalogging Thermo Hygrometer 800020 Instruction Manual ...

Page 2: ...DUCTION 3 FEATURES 4 MATERIALS SUPPLIED 4 FRONT PANEL DESCRIPTION 5 LCD DISPLAY 6 SET UP 7 MEASUREMENT PROCEDURES 12 SOFTWARE INSTALLATION 17 DATALOGGING 19 POWER SUPPLY 24 CARE AND MAINTENANCE 25 SPECIFICATIONS 26 WARRANTY 28 ...

Page 3: ...meter and later downloaded or conveniently e mailed if a PC is unavailable at the time All data can be viewed and stored in Excel format or you can use our unique graphing tools in the included software to view and analyze the data Calibrate to known standards using the offset function for humidity ambient air temperature and ice point The meter comes ready to use with a protective foam lined carr...

Page 4: ...nts Direct upload of data to EXCEL text or graph files Customizable data graphing Automatic shutoff Backlight Bluetooth streaming and data sharing MATERIALS SUPPLIED Datalogging Thermometer Micro USB Cable One K type Beaded Wire Probe Free Software Download at www sperdirect com software htm Free App Download at Instruction Manual Four AAA batteries Carrying Case Bluetooth 7 pin adapter ...

Page 5: ...on 4 LEFT Arrow and MIN MAX 5 RECALL Memory Button 6 SET Button 7 Backlight Button 8 Up Arrow and Hold Button 9 Right Arrow and UNIT Selection 10 Down Arrow and REC Button 11 MEM Button 12 Thermocouple Input 13 Bluetooth Chip 14 Micro USB Port 15 Bluetooth door 16 Tilt Stand 17 Battery Chamber ...

Page 6: ...ed 2 Bluetooth Activated 3 Wet Bulb Active on T1 4 Current RH Reading 5 Auto Power off is Enabled 6 Low Battery Indicator 7 Min Max Average Activated 8 Dew Point Active on T1 9 Current Ambient Air T1 Reading 10 Current Thermocouple Reading T2 ...

Page 7: ...tup Mode The meter will return to Normal Mode if the SET button is pressed Bluetooth Selection 1 Press SET to enter Setup Mode 2 bluetooth appears on the display along with the word on or off Press or to select on 3 Press MEM to save the selection and toggle to the next setting Otherwise press SET to exit setup mode Thermocouple Type 1 Press SET to enter Setup Mode 2 Press MEM to toggle through th...

Page 8: ...ugh the settings until Int appears on the display along with the time interval in minutes and seconds The minutes flash on the display 3 Press or to adjust the minutes 4 Press or to move the cursor to the required number of seconds The seconds will flash 5 Press or to adjust the seconds 6 Press MEM to save the selection and toggle to the next setting Otherwise press SET to exit setup mode Offset V...

Page 9: ...Use the or to advance through the placeholders and the or buttons to change the numbers Press MEM to save the selection and toggle to the next parameter Note The alarm function is either on for all parameters or off for all parameters The best way to alarm only one parameter is to set the values of the unwanted parameters to something not possible during your measurement period For example if you ...

Page 10: ... 30 minutes or hour increments of 1 2 4 or 8 5 Press MEM to save the selection and toggle to the next setting Otherwise press SET to exit setup mode Setting the Date and Time 1 Press SET to enter setup mode 2 Press MEM to toggle through the settings until the Year Date and time appear on the screen along with the word SET 3 Press or to adjust the year 4 Use the or to advance to the date 5 Press or...

Page 11: ... cover 3 Place the Bluetooth adapter into the housing with the label facing up 4 Slide the adapter up to connect the pins of the adapter with the pins of the meter 5 Screw the plastic cover back on to the meter 6 The meter is now Bluetooth capable Please Note that While most Smartphones have built in Bluetooth capability many PC s DO NOT ...

Page 12: ...n To toggle between them press DP WB Data Hold Note The MIN MAX AVG DP WB RECALL and SETUP buttons on the meter are disabled during data hold 1 Press HOLD to freeze the reading on the display Hold appears at the top of the LCD 2 Press HOLD to return to Normal Mode Maximum Minimum Average Mode Note Viewing the maximum minimum and average values will not interrupt temperature measurement The maximum...

Page 13: ...he saved reading from 00 to 99 To recall a reading 1 Press RECALL The RECALL symbol lights up on the display along with the associated serial number 2 Use the or buttons to select a saved memory serial number The LCD shows the time of the reading for 2 seconds followed by the saved data 3 Press RECALL to return to Normal Mode Clearing Readings in Memory 1 Press and hold POWER for 3 seconds to turn...

Page 14: ...trieving it Option 1 Download the data to a smartphone by following the instructions on page 15 Option 2 Connect the meter to a PC using the included micro USB cable Retrieve the data using the PC software as described on page 23 Note the meter will not record if battery power is extremely low Clearing the Recorded Data 1 Press and hold Power for 3 seconds to turn the meter off 2 Press and hold RE...

Page 15: ... instructions for doing this are on page 7 of this manual 4 Pair the meter with the cell phone s Bluetooth under the settings menu on the phone Since the phone can remember several meters it is important to note that the meter s name will match the serial number of the Bluetooth adapter installed inside the meter 5 Open the application on the phone If the meter is not yet paired the application wi...

Page 16: ...d you will see real time readings graphs and download data from the meter by simply clicking the icons for each of these features 7 Data can be saved retitled e mailed and sent via text message from the application See image below ...

Page 17: ...the Sper Scientific website at www sperdirect com software htm If Windows does not run the Setup exe automatically open the download folder and double click the file from there 2 Select Yes to allow the program Setup exe to make changes to your computer 3 Select Next to start the installation 4 Select the folder to install the files and click Next 5 Select the location for the program shortcuts an...

Page 18: ...s 1 Click on the Start Menu 2 Select Control Panel 3 Launch the Add Remove Programs 4 Highlight 800020 5 Click on Add Remove 6 Click Next 7 Click Next 8 Click Finish OR 1 Click on the Start Menu 2 Select Control Panel 3 Select Programs and Features 4 Double click on 800020 5 The Installation Dialog Box appears Select Uninstall and click Next 6 Click Next 7 Click Finish ...

Page 19: ... some PC s but is usually done through the settings menu under Add a Device On some computers the device may be recognized with both a headphone icon and a keyboard icon see image below If this is the case always choose the headphone icon 3 Insert the thermocouple s into the appropriate port s 4 Open the Sper Scientific software An image of the meter displays along with a graph ...

Page 20: ...he software to the meter by clicking the Bluetooth icon see the image below Please note that the meter s Bluetooth must be on to connect to the software Do this by following the instructions on page 7 of this manual ...

Page 21: ...ck anywhere on the graph to mark it with an X This option is not available when Split is selected Change cursor to an I Click anywhere on the graph to add an annotation This option is not available when Split is selected Set annotation color ensuring sufficient contrast with the graph background Split Separate or combine the four parameters in the graph When Split is not selected the graph will us...

Page 22: ...ort the graph to the clipboard a file or a printer 2 Click Font to select the font name style and size for annotations 3 Click Y Axis to customize the maximum and minimum values on the graph Saving Real Time Data 1 Click the graph window you want to save The selected graph window will become active 2 Click File and select Save or simply click the disc icon 3 Enter the file name and file type exten...

Page 23: ...Recorded Data Select Data Logger from the main menu bar or click the datalogger icon under the main menu An indicator shows the loading progress The left side of the screen will display the number of data sets with detail information for each set The first data set will appear in the graph on the right side Click any data set to graph that set ...

Page 24: ...crew and remove the screw at the bottom of the battery cover 2 Lift off the battery cover 3 Insert four new AAA batteries ensuring correct polarity 4 Replace the battery cover and reinstall the screw Replace the batteries when the low battery icon blinks on the LCD Note Before replacing the batteries turn the meter off and disconnect the temperature probes ...

Page 25: ...ee antistatic cloth Do not use abrasives solvents or cleaning agents containing carbon alcohol or benzenes on the meter Repairs or services not covered in this manual should be performed by qualified personnel only Please contact Sper Scientific to speak with a technician ...

Page 26: ...2 F 0 to 1767 C 0 5 F 1000 F 1 F 1000 F 0 2 C 600 C 1 C 600 C 0 2 RDG 2 5 F 0 2 RDG 1 4 C Temperature Coefficient 0 01 RDG 0 0028 F 0 05 C per C 64 F 18 C or 82 F 28 C Sample Rate 2 times per second Battery Type UM 4 or AAA 1 5V battery x 4 Battery Lifetime Bluetooth OFF Approx 100 hours alkaline battery Bluetooth ON Approx 20 hours alkaline battery Operating Temperature 32 to 122 F 0 to 50 C Oper...

Page 27: ...bulb 4 to 140 F 20 to 60 C 0 1 F 0 1 C 4 to 140 F 20 to 60 C 1 5 F 0 8 C Dew point 58 to 140 F 50 to 60 C 0 1 F 0 1 C Temperature 4 to 140 F 20 to 60 C 0 1 F 0 1 C Response Time t90 Humidity 60 sec in slowly moving air Temperature 20 sec in slow moving air T1 Sensor hysteresis 1 5 RH from 10 to 90 Temperature coefficient 0 1 x specified accuracy per 1 C 73 4 F 23 C or 73 4 F 23 C ...

Page 28: ... This warranty does not cover probes batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will void the warranty To obtain warranty service ship the unit postage prepaid to SPER SCIENTIFIC LTD 8281 E Evans Rd Suite 103 Scottsdale AZ 85260 480 948 4448 The defective unit must be accompanied by a description ...

Reviews: