8
Unit of Measure
1. Press
SET
to enter setup mode.
2. Press
MEM
to toggle through the settings until “Unit”
appears on the display along with flashing symbols for
°C, °F, or K. Press ▲or ▼ to cycle to your desired
temperature unit.
3. Press
MEM
to save the selection and toggle to the next
setting. Otherwise, press
SET
to exit setup mode.
Time Interval
The time interval range for data storing is 1 second to 60
minutes 59 seconds. The time interval defines how often a
reading is taken during record mode.
1. Press
SET
to enter setup mode.
2. Press
MEM
to toggle through the settings until “Int”
appears on the display along with the time interval in
minutes and seconds.The minutes flash on the display.
3. Press ▲or ▼ to adjust the minutes.
4. Press ◄ or ►to move the cursor to the required number
of seconds. The seconds will flash.
5. Press ▲or ▼ to adjust the seconds.
6. Press
MEM
to save the selection and toggle to the next
setting. Otherwise, press
SET
to exit setup mode.
Offset Values
All parameters can be adjusted to compensate for probe
errors. The offset range is ± 5°C or ± 9°F.
1. Press
SET
to enter setup mode.
2. Press
MEM
to toggle through the settings until OFFSET
appears at the bottom of the screen. Press ▲or ▼to
increase or decrease the offset value as compared to
a known standard such as an ice bath or humidity salt
chamber.