DF3/FF3 MANUAL
13
118018-001
Rev E
3.2.2 REMOTE
MONITOR:
Test points are made available at the rear 50-pin
connector from monitoring tube voltage, tube
current and tube filament current. Also, tube
voltage programming and tube current
programming readback signals are provided. The
test points are always positive regardless of the
output polarity, where 0 to 10V = 0-100% of
output. Accuracy is within
0.5%. All test points
have output impedance at 1Kohm. See Figure
3.3 for remote monitor interface.
38
47
JB1
M
KV Monitor
Mon Common
2) All signals 0 - 10Vdc = 0 - 100% of specified maximum outputs.
1) Metering Circuit should be high Z type (>10 Meg) for accurate measurement.
NOTES:
Customer Interface
High Voltage Power Supply
1K
mA Monitor
1K
39
M
Monitor
1K
46
M
PS Common
Filament
Figure 3.3 REMOTE MONITOR