Choosing a Mounting Location

When considering a location for the installation of the 
RO System, consider the following factors:

Light Sources

1.    Most of the components of this system are plastic and 
 

 are subject to damage by ultraviolet light from the sun 

 

 and other sources such as metal halide  lighting.

2.   Algae is more likely to thrive inside the clear filter housings 
 

 when exposed to bright light.

3.   Avoid installing this unit in bright light or direct sunlight.

Temperature Extremes

1.   The unit must be kept out of areas that are subject to freezing
  

temperatures.

2.   High temperatures greater than 100° F (38° C) must be 
 

 avoided. If the unit is used outside, avoid putting the 

 

 system in direct sunlight or connecting it to a

 

 garden hose that may be exposed to sunlight.

System Care

SpectraPure

®

SpectraPure®Inc

 Fax  480.894.6109 Fax us toll-free 1.877.527.7873

E-mail: [email protected] Visit us on the web www.spectrapure.com

3

System Information

System Specifications

Sediment Pre-Filter 

 

1 micron MicroTec™  sediment pre-filter

Carbon Filter  

 

 

10 micron carbon block pre-filter

RO Membrane Type   

Thin-Film Composite  (TFC)

DI Cartridges  

 

 

SilicaBuster™ Color-Indicating DI Cartridge

Rejection Rate   

 

Greater than 98% average

Input Water Pressure 

60 psi (4.15 bar) line pressure*

Input Water Temp    

77°F (25°C)

Recovery Rate   

 

20% (i.e. 20% of the water will be

       collected 

as 

pure 

water)

Nominal Membrane Flow Rates @ 60 psi  & 77° F :
 

GPD (lpd) 

Product Water Flow Rate   

Concentrate Flow Rate

 

  25 (95)   

 

 66 ml/min   

 

 

 

264 ml/min

  

 

40 

(151)  

 105ml/min 

    420 

ml/min

 

  60 (228) 

  

158 ml/min   

 

 

 

632 ml/min

  

 

90 

(340)  

 235ml/min 

    940 

ml/min

 

180 (680) 

 

470ml/min   

 

 

      1880 ml/min

Reverse Osmosis Feed Water Requirements

Operating Pressure*   

 

 

40 – 80 psi (2.75 – 5.5 bar)

pH 

Range      3 

– 

11

Maximum Temperature 

 

 

100° F (38° C)

Maximum 

Turbidity    1.0 

NTU

Maximum Silt Density Index 

 

5.0 (based on 15 min. test time)

Maximum Chlorine 

 

 

 

less than 0.1 ppm

Maximum 

TDS     2000 

ppm

Maximum Hardness   

 

 

10 grains (170 ppm as CaCO3)

Maximum 

Iron     less 

than 

0.1 

ppm

Maximum Manganese  

 

 

less than 0.1 ppm

Maximum Hydrogen Sulfide 

 

0 ppm

Langlier Saturation Index   

 

LSI must be negative

*Operating pressure less than 40 psi will require a booster pump:
 

less than 49 GPD use BPLF-MO-115(-230),

 

more than 49 GPD use BPHF-MO-115(-230).

*Operating pressure greater than 80 psi will require a pressure reducing valve.

Summary of Contents for MaxPure MPDI RO/DI System

Page 1: ...performance of membrane elements for one year from date of receipt by the buyer providing that the loss of performance was not caused by fouling neglect or water conditions exceeding the feed water parameters listed on page 3 of this manual Most municipal water supplies meet these requirements Terms and Conditions of Sale 1 To obtain service under this warranty the defective system or components m...

Page 2: ...echanical photocopying recording or otherwise without the prior written permission of SpectraPure Inc SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 23 Replacement Parts Model Replacement Part SF MT 1 0 10 1 micron MicroTec Sediment Filter CF 10 10 10 micron Carbon Block Pre Filter SF MT 0 5 10 0 5...

Page 3: ...s Sediment Pre Filter 1 micron MicroTec sediment pre filter Carbon Filter 10 micron carbon block pre filter RO Membrane Type Thin Film Composite TFC DI Cartridges SilicaBuster Color Indicating DI Cartridge Rejection Rate Greater than 98 average Input Water Pressure 60 psi 4 15 bar line pressure Input Water Temp 77 F 25 C Recovery Rate 20 i e 20 of the water will be collected as pure water Nominal ...

Page 4: ...oon as the cartridge becomes exhausted Operating the Push Fittings The push fittings used on SpectraPure products can be easily serviced To remove the tubing from its push fitting a Firmly depress and hold the push fitting collar down with your thumbnail b While the push fitting collar is depressed pull the tubing straight out of the push fitting Once the tubing is removed release the collar Syste...

Page 5: ...e Membrane Test Kit RO product water Y greater than Yes Membrane OK for Alkalinity 90 Hardness Measure feed water X Is Rejection No Replace Membrane Test Kit RO product water Y greater than Yes Membrane OK for Hardness 90 Caution This test is not to be used on softened water sources Membrane Testing SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectra...

Page 6: ...he RO membrane SpectraPure Inc assumes no responsibility for water damage due to leaks It is the user s responsibility to determine that the system is leak free SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 19 Membrane Output Calculation Example What is the expected GPD from a 75 GPD System at 40 ...

Page 7: ...ectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 7 System Set Up Checking the Concentrate to Purified Water Ratio This procedure will assure you of maximum life and reliability of your SpectraPure System Failure to perform this procedure can permanently damage the membranes and will void the pro rated ...

Page 8: ...rt into the end of the yellow tubing by carefully pressing on the insert with your thumbnail Care should be taken not to crush or other wise damage the end of the capillary tubing protruding from the end of the insert SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 17 Deionization Cartridge Diagnost...

Page 9: ...duct water tube double O ring end of membrane product water stem tube Diagnostics System Maintenance SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 9 System Set Up Fig B Reverse Osmosis Assembly Top Rear View membrane feed push fitting black tubing to membrane feed RO membrane yellow concentrate tu...

Page 10: ... Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 15 RO Membrane Replacement 1 Turn off the water supply to the RO system Place the system where the membrane housing is easily accessible 2 Remove the black tubing from the membrane feed push fitting by depressing the collar on the fitting with your thumb and pulling the tubi...

Page 11: ...Procedure 1 Turn on the system let it operate long enough to fill the empty DI housing then let it run for an additional 20 minutes 2 See the section on Testing the Quality of the Membrane page 19 Perform the test using your chosen method 3 Turn off the system drain the DI housing re install the DI cartridge and turn on the system Note All water sources are different and are subject to changes in ...

Page 12: ...ectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 13 Carbon Block Filter Diagnostic Replace the Carbon Block Filter at least every 6 months OR when chlorine breakthrough greater than 0 1 ppm occurs in the yellow con centrate line To test for chlorine breakthrough collect a 10 ml sample of the concentrat...

Reviews: