background image

SpectraPure

SpectraPure®Inc.

  

480.894.5437   Call us toll-free  1.800.685.2783    

215 South Industrial Drive,   Suite 2A   Tempe, Arizona   85281-2941

®

6

1.  Remove the DI cartridge from the right-most housing and 
 

re-install the empty housing.

2.  Attach the garden hose adapter to your cold water source. 
 

Never run hot water (greater than 100° F (38° C)) through

 the 

system.

3.  Place the yellow concentrate tubing and the blue purified 
 

water tubing into the drain. Do not restrict flow from these 

 lines.
4.  Slowly open the cold water supply valve and allow 
 

the first two housings to fill. You may use pressure up to 

 

80 psi (5.5 bar). 

5.  Check the system to ensure that all fittings are tight and 
 

leak-free before leaving the system unattended.

6.  Allow the system to produce at least 2 gallons (7.57 liters) 
 

of purified water and discard.

7.  Check the Concentrate to Purified Water Ratio by follow
 

ing the set-up instructions on the next page.

8.  Close the cold water supply valve. Locate and re-install the 
 DI 

cartridge.

9.  Slowly open the cold water supply valve and discard the
 

first gallon of product water.

Note:  Air trapped in the DI cartridges is a normal condition 
and will not affect the operation of the DI cartridges. Mounting 
the blue product water tubing so that at least a short length 
of tubing is higher than the filter unit can often alleviate this 
condition.

System Start-Up

System Initialization & System Start-Up

If you are setting up your system for the first time or replacing 
the RO membrane:

SpectraPure®Inc

.

 assumes no responsibility for water 

damage due to leaks. It is the user’s responsibility to determine

that the system is leak-free.

SpectraPure

®

SpectraPure®Inc

 Fax  480.894.6109 Fax us toll-free 1.877.527.7873

E-mail: [email protected] Visit us on the web www.spectrapure.com

19

Membrane Output Calculation Example

What is the expected GPD from a 75 GPD System at 40 psi pressure and 60°F 
water temperature?
 

PCF = 40 ÷ 60 = 0.666

 

TCF = 0.754 (from Table 1)

 

Expected GPD = 75 × 0.666 × 0.754 = 37.7 GPD ± 15%

 

37.7 GPD would be the Actual Production Rate

Testing the Quality of the Membrane

The performance of a RO membrane is measured by its ability to reject salts 
(or TDS (Total Dissolved Solids)). 

Important: Test the quality of the membrane once every 6 months.

Note: This procedure will require a Conductivity Meter (TS-C61) or (TS-T71).

Procedure:

 

1.  Measure tap water conductivity. (Call it X)

 

2.  Run the system for 15-20 minutes.

 

3.  Rinse test instrument cell 2-3 times with RO water.

 

4.  Measure RO water conductivity directly from the blue product 

 

 

water line. (Call it Y).

 

5.  Subtract RO water conductivity from tap water conductivity. (X - Y) 

 

6.  Divide this quantity by tap water conductivity. (X - Y) ÷ X 

 

7.  Rejection = [(X - Y) ÷ X ] ×100 

* Conductivity in the above procedure could be replaced by hardness, alkalinity, 
nitrate, phosphate, silica etc. (measured in ppm or mg/l).

Rejection of the RO Membrane Calculation Example

 

1.  Tap water hardness = 150 ppm (X)

 

2.  RO water hardness = 7 ppm (Y)

 

3.  X - Y = 143 ppm

 

4.  (X - Y) ÷ X = 143 ÷ 150 = 0.953

 

5.  Rejection = [ ( X - Y)  ÷ X ] ×100 = 0.953 ×100 = 95.3

 

Membrane Hardness Rejection = 95.3 % :  Rejection rates less than 95% may

 

 indicate that the membrane should be replaced.

Membrane Testing

Summary of Contents for MaxPure MPDI RO/DI System

Page 1: ...performance of membrane elements for one year from date of receipt by the buyer providing that the loss of performance was not caused by fouling neglect or water conditions exceeding the feed water parameters listed on page 3 of this manual Most municipal water supplies meet these requirements Terms and Conditions of Sale 1 To obtain service under this warranty the defective system or components m...

Page 2: ...echanical photocopying recording or otherwise without the prior written permission of SpectraPure Inc SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 23 Replacement Parts Model Replacement Part SF MT 1 0 10 1 micron MicroTec Sediment Filter CF 10 10 10 micron Carbon Block Pre Filter SF MT 0 5 10 0 5...

Page 3: ...s Sediment Pre Filter 1 micron MicroTec sediment pre filter Carbon Filter 10 micron carbon block pre filter RO Membrane Type Thin Film Composite TFC DI Cartridges SilicaBuster Color Indicating DI Cartridge Rejection Rate Greater than 98 average Input Water Pressure 60 psi 4 15 bar line pressure Input Water Temp 77 F 25 C Recovery Rate 20 i e 20 of the water will be collected as pure water Nominal ...

Page 4: ...oon as the cartridge becomes exhausted Operating the Push Fittings The push fittings used on SpectraPure products can be easily serviced To remove the tubing from its push fitting a Firmly depress and hold the push fitting collar down with your thumbnail b While the push fitting collar is depressed pull the tubing straight out of the push fitting Once the tubing is removed release the collar Syste...

Page 5: ...e Membrane Test Kit RO product water Y greater than Yes Membrane OK for Alkalinity 90 Hardness Measure feed water X Is Rejection No Replace Membrane Test Kit RO product water Y greater than Yes Membrane OK for Hardness 90 Caution This test is not to be used on softened water sources Membrane Testing SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectra...

Page 6: ...he RO membrane SpectraPure Inc assumes no responsibility for water damage due to leaks It is the user s responsibility to determine that the system is leak free SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 19 Membrane Output Calculation Example What is the expected GPD from a 75 GPD System at 40 ...

Page 7: ...ectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 7 System Set Up Checking the Concentrate to Purified Water Ratio This procedure will assure you of maximum life and reliability of your SpectraPure System Failure to perform this procedure can permanently damage the membranes and will void the pro rated ...

Page 8: ...rt into the end of the yellow tubing by carefully pressing on the insert with your thumbnail Care should be taken not to crush or other wise damage the end of the capillary tubing protruding from the end of the insert SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 17 Deionization Cartridge Diagnost...

Page 9: ...duct water tube double O ring end of membrane product water stem tube Diagnostics System Maintenance SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 9 System Set Up Fig B Reverse Osmosis Assembly Top Rear View membrane feed push fitting black tubing to membrane feed RO membrane yellow concentrate tu...

Page 10: ... Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 15 RO Membrane Replacement 1 Turn off the water supply to the RO system Place the system where the membrane housing is easily accessible 2 Remove the black tubing from the membrane feed push fitting by depressing the collar on the fitting with your thumb and pulling the tubi...

Page 11: ...Procedure 1 Turn on the system let it operate long enough to fill the empty DI housing then let it run for an additional 20 minutes 2 See the section on Testing the Quality of the Membrane page 19 Perform the test using your chosen method 3 Turn off the system drain the DI housing re install the DI cartridge and turn on the system Note All water sources are different and are subject to changes in ...

Page 12: ...ectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 13 Carbon Block Filter Diagnostic Replace the Carbon Block Filter at least every 6 months OR when chlorine breakthrough greater than 0 1 ppm occurs in the yellow con centrate line To test for chlorine breakthrough collect a 10 ml sample of the concentrat...

Reviews: