![Spectranetics CVX-300 Operator'S Manual Download Page 42](http://html.mh-extra.com/html/spectranetics/cvx-300/cvx-300_operators-manual_1343921042.webp)
42
7030-0068-26
28Jul16
Customer Service, Canada/US (719) 633-8333 / (800) 231-0978
Customer Service, 31 33 434 7050
CVX-300
®
Excimer Laser System
Version 26
Operator’s Manual
English / English
7.
While the Excimer Laser System is lasing, the total treatment time is stored. To retrieve
the treatment time, stop lasing and depress the Treatment Time button. The Min/Sec
indicator light illuminates and the total treatment time is visible in the display window.
(Figure 11)
8.
To reset the treatment time counter, depress the Reset and Treatment Time buttons
simultaneously.
9.
While the Excimer Laser System is lasing, the number of pulses delivered during a
procedure is accumulated. To retrieve the total number of pulses delivered, stop lasing
and depress the Pulses Delivered button. The count is visible in the display window.
(Figure 12)
10.
To reset the pulses delivered counter, depress the Reset and Pulses Delivered buttons
simultaneously.
Figure 11
Treatment Time