Advisory, Indicator and Status Lights
Aim Fiber Energy and Millijoules: The Aim Fiber advisory light and the Energy and
Millijoules indicator lights illuminate simultaneously indicating the system is in the
Calibrate mode or Read Energy has been selected in the Ready mode. To begin calibration,
aim the fiber tip at the calibration detector head and depress the footswitch. The system
reads and displays the fiber output energy after each laser pulse.
Calibrate Button: The Calibrate button illuminates indicating the Excimer Laser System is
operating in Calibrate mode.
Calibrate Light: The Calibrate advisory light indicates the Excimer Laser System needs to
be calibrated.
Cal OK: The Cal OK advisory light indicates the Excimer Laser System has successfully
completed calibration.
YELLOW
when Power
Error lamp
is ON
GREEN
when in
STANDBY
mode
YELLOW
when in
READY
mode
RED when
LASING
Status lamps
at top of
control panel