Figure 8
Standby
Standby Mode (Returning from Ready Mode)
1.
Standby mode is entered from Ready mode by depressing the Standby button.
2.
The Standby button and green status light on the top of the control panel
are illuminated.
3.
The fluence value and pulse repetition rate appear in the display window.
4.
The Cal OK advisory light remains illuminated. (Figure 8)
Ready Mode
1.
Ready mode is entered from Standby mode by depressing the Ready button. (Figure
9)
2.
The Ready button, Cal OK light and yellow status light on the top of the control panel
are illuminated.
3.
The fluence value and pulse repetition rate may be changed by depressing a Fluence
or Rate button. After a change to either value, it is not necessary to re‑calibrate.