Install the replacement battery into the holder and reinstall the holder cover.
Reconnect the two-position plug to the front panel circuit board. Verify the plug is
properly mated to the header to maintain the correct polarity.
Reinstall the top cover.
Reapply power to the unit.
NOTE:
The Replace Battery lamp will remain on after replacing the battery. The lamp
can be reset using the RS-232 command CA BAT, or will self-clear after
successfully completing the first battery test (typically 48 hours).
Connect the PC to the front panel RS-232 Com port and issue the command
CA BAT
to
reset the latched Replace Battery alarm lamp.
The Reset Battery Alarm command is a Set Mode command.
Place the unit in Set Mode by sending the following command:
Type:
SM ON <ent>
The unit will respond with an acknowledgement that Set Mode is enabled. Issue
Battery Reset command as follows:
Type:
CA BAT <ent>
The Battery Status lamps shall now indicate that the battery is Charging or is
Ready.
This completes the battery replacement procedure. The unit will require a two to three
hour period to obtain oscillator lock to the GPS reference.