background image

 

11 

 

(a) 

 (b) 

    

 

Fig.3.4.  Displaying  information  on  the  IN-01m  stress 

indicator screen when the power is turned on (a) and after the 
measurement mode is set (b) 

3.4.3.  Button  6  is  used  to  turn  on  the  stress  indicator 

screen  backlight.  When  you  turn  on  button  6,  the  screen  is 
illuminated  with  additional  light  to  improve  the  display  of 
digital stress indicator readouts. 

3.4.4.  Button  7  is  used  to  save  measurements  to  the 

electronic memory of the stress indicator. When the button 7 is 
turned  on  or  off,  the  current  transmitter  reading  displayed  on 
the screen is recorded in the stress indicator memory. 

3.4.5.  When  connecting  a  cable  connected  to the USB 

port  of  a  personal  computer  to  the  mini-USB  11  connector, 
information about establishing  a connection  with  the personal 
computer appears on the stress indicator screen (Fig. 3.5). 

 

Fig.3.5.  Displaying  information  on  the  IN-01m  stress 

indicator screen when connected to a personal computer 

 

Using of the IN-01m Connect software application pre-

installed  on  a  personal  computer,  the  measured  information 
from  the  stress  indicator's  electronic  memory  is  transmitted 
over a cable and stored in the personal computer's memory. 

3.4.6.  Button  8  displays  the  current  date  and  time  of 

stress indicator on the screen. 

 
 

 

12 

 

3.5. Magnetizing unit 

3.5.1.  The  magnetizing  unit  of  the  stress  indicator  is 

used for magnetizing the local metal section of a steel product 
or metal structure with a constant magnetic field, the intensity 
vector of which is directed perpendicular to the metal surface. 
 

3.5.2.  The  magnetizing  device  is  non-volatile,  single-

pole, rod type. It uses permanent magnets made from rare earth 
metals. 
 

3.5.3.  The  pole  of  the  magnetizing  device  is  protected 

by an overlay to ensure convenience in the process, the handle 
of the magnetizing device is provided with a strap. 
 

 

3.6. Sensor 

3.6.1.  The  stress  indicator  sensor  is  designed  to 

measure  the  field  strength  of  the  residual  magnetization  of 
metal over the  surface of a  local  section of a  steel product or 
metal structure. 

3.6.2.  The  sensor  contains  a  precision  integrated  Hall 

Effect  sensor  chip  and  a  precision  integrated  temperature 
sensor chip with an analog signal output, which are located in 
a single housing. 

3.6.3. The Hall  sensor plane  is  located in the center of 

the  transducer  housing  parallel  to  the  working  surface  at  a 
distance of 0.5 mm from it.  

3.6.4.  To  measure  the  field  strength  of  the  residual 

magnetization  of  the  metal,  the  center  of  the  working  surface 
of the stress indicator 

sensor

 is set in the middle of the residual 

magnetized area of the metal. 

 

 

 
 
 
 
 

Summary of Contents for IN-01m

Page 1: ...Spector Sci Tech LLC IN 01m METAL STRESS INDICATOR USER MANUAL Russia Ufa 2020 2...

Page 2: ...1 PURPOSE 1 1 The IN 01m metal mechanical stress indicator is intended for indirect estimation of the value of the metal mechanical stress of a local section of a steel product or a metal structure ma...

Page 3: ...rent consumed from a freshly charged PP3 element mA when working without the display backlight 7 0 5 when working with the display backlight 16 0 5 2 6 Operating time of stress indicator when powered...

Page 4: ...effect sensor 2 via a scale amplifier combined with a 4 6 7 5 3 2 1 8 8 low pass filter 4 to the input of the signal level exceeding light indication unit 5 and the input of the analog to digital conv...

Page 5: ...receiving service information from the computer 3 3 4 4 Connector 12 is used to connect the charger s network adapter to the electronic unit for recharging the battery battery 3 4 The purpose of the...

Page 6: ...lays the current date and time of stress indicator on the screen 12 3 5 Magnetizing unit 3 5 1 The magnetizing unit of the stress indicator is used for magnetizing the local metal section of a steel p...

Page 7: ...rmation first appears then the value of the measured signal of the Sensor is set see section 3 4 2 of current document 14 5 1 5 In low light conditions to improve the visibility of the stress indicato...

Page 8: ...ation according to step 6 2 1 6 2 2 Install the stress indicator sensor in the middle area of the magnetized end surface of the steel stud or bolt 6 2 3 To measure the field strength of the residual m...

Page 9: ...me N Measurement N End of file 18 The data file represents sequentially recorded measurement results see table 6 1 6 4 3 Transfer of data recorded in the stress indicator s electronic memory to a pers...

Page 10: ...In the event of failure of the electronic components of the stress indicator they must be replaced 7 5 Current repairs are made during the operation of the stress indicator At the same time faults de...

Page 11: ...must ensure their stable position exclude the possibility of collision with other objects as well as against the walls of the vehicle 9 5 Conditions of transportation temperature C from minus 30 to p...

Page 12: ...11 0 11 5 12 14 5 650 400 5 6 8 0 8 5 10 11 Table A 2 r mT Steel grade B MPa Y MPa 0 Y B 3 350 210 0 62 1 69 2 0 09 2 470 325 1 06 1 38 2 43 2 79 2 98 17 1 520 350 1 38 3 12 4 12 Steel 20 420 230 1 3...

Reviews: