background image

 

1.8.  According  to  the  operating  conditions,  the  stress 

indicator  belongs  to  the  type  of  climatic  version  of  UHL  3.1 
according to GOST 15150-69 and can work steadily when: 

- ambient temperature from minus 20ºC to plus 70ºC, 

- relative humidity up to 98 % at +25ºC, 

- atmospheric pressure from 84 kPa to 106.7 kPa. 
 
 

2. TECHNICAL SPECIFICATIONS 

 

2.1. The range of  measurement of the field  strength of 

the residual magnetization of the metal, mT: 

- from 0 to 20.00

 

2.2.  Measuring  range  of  mechanical  stress  of  metal  of 

steel products, metal structures, MPa: 

- from 0 to σ

B

  

where σ

B

 is the ultimate strength of the metal. 

 

2.3.  The  limit  of  the  allowed  basic  error  in  measuring 

the  mechanical  stress  of  the  metal  of  a  steel  product or  metal 
structure is determined by the formula: 

0.5 + 0.05 σ

 

(2.1) 

where σ

i

 is the stress indicator readout in MPa. 

 

2.4.  The  metal  stress  indicator  is  powered  by  a  PP3 

element with nominal electric voltage of 9V. 

 

2.5.  Electric  current  consumed  from  a  freshly  charged 

PP3 element, mA: 

- when working without the display backlight 

- 7 ± 0.5

 

- when working with the display backlight 

- 16 ± 0.5 

2.6.  Operating  time  of  stress  indicator  when  powered 

from freshly charged element PP3, in hours, at least: 

- 20 

 

2.7.  Time  to  set  the  metal  stress  indicator's  operating 

mode, s, not more than: 

- 10 

2.8. Time of one measurement, s, not more than: 

- 5 

2.9.  Time  of  continuous  operation  of  the  stress 

indicator, h, not less than:  

- 16 

2.10. Battery charge time, minutes, not more: 

- 80 

2.11.  The  stress  indicator  has  a  memory  card  with  a 

capacity of more than 60000 measurements.

 

 

2.12.  The  stress  indicator  has  a  standard  output  for 

transmitting  measured  information to a personal computer via 
a mini-USB connector. 

2.13. Overall dimensions, mm, not more than: 

- Electronic unit (length × width × thickness) 

 

- 136×72×28

 

- Sensor (diameter × length) 

- 20×15

 

- Sensor cable length 

-950±50 

- Magnetizing device (diameter × length) 

- 20×190 

2.14. Weight, g: 

- Electronic unit with battery 

- 200 

- Sensor with cable 

- 40 

- Magnetizing device 

- 300 

2.15. Average recovery time, h:  

- 5 

2.16. Full average service life, years: 

- 10 

2.17. Established service life, years: 

- 2 

 

 

 
 

Summary of Contents for IN-01m

Page 1: ...Spector Sci Tech LLC IN 01m METAL STRESS INDICATOR USER MANUAL Russia Ufa 2020 2...

Page 2: ...1 PURPOSE 1 1 The IN 01m metal mechanical stress indicator is intended for indirect estimation of the value of the metal mechanical stress of a local section of a steel product or a metal structure ma...

Page 3: ...rent consumed from a freshly charged PP3 element mA when working without the display backlight 7 0 5 when working with the display backlight 16 0 5 2 6 Operating time of stress indicator when powered...

Page 4: ...effect sensor 2 via a scale amplifier combined with a 4 6 7 5 3 2 1 8 8 low pass filter 4 to the input of the signal level exceeding light indication unit 5 and the input of the analog to digital conv...

Page 5: ...receiving service information from the computer 3 3 4 4 Connector 12 is used to connect the charger s network adapter to the electronic unit for recharging the battery battery 3 4 The purpose of the...

Page 6: ...lays the current date and time of stress indicator on the screen 12 3 5 Magnetizing unit 3 5 1 The magnetizing unit of the stress indicator is used for magnetizing the local metal section of a steel p...

Page 7: ...rmation first appears then the value of the measured signal of the Sensor is set see section 3 4 2 of current document 14 5 1 5 In low light conditions to improve the visibility of the stress indicato...

Page 8: ...ation according to step 6 2 1 6 2 2 Install the stress indicator sensor in the middle area of the magnetized end surface of the steel stud or bolt 6 2 3 To measure the field strength of the residual m...

Page 9: ...me N Measurement N End of file 18 The data file represents sequentially recorded measurement results see table 6 1 6 4 3 Transfer of data recorded in the stress indicator s electronic memory to a pers...

Page 10: ...In the event of failure of the electronic components of the stress indicator they must be replaced 7 5 Current repairs are made during the operation of the stress indicator At the same time faults de...

Page 11: ...must ensure their stable position exclude the possibility of collision with other objects as well as against the walls of the vehicle 9 5 Conditions of transportation temperature C from minus 30 to p...

Page 12: ...11 0 11 5 12 14 5 650 400 5 6 8 0 8 5 10 11 Table A 2 r mT Steel grade B MPa Y MPa 0 Y B 3 350 210 0 62 1 69 2 0 09 2 470 325 1 06 1 38 2 43 2 79 2 98 17 1 520 350 1 38 3 12 4 12 Steel 20 420 230 1 3...

Reviews: