background image

 

CONTENTS 

 

 

 

 

 

 

 

Page 

1. Purpose .............................................................................. 4 

2. Technical specifications ..................................................... 5 

3. Device and principle of operation....................................... 7 

4. Safety instructions............................................................ 13 

5. Preparing the metal stress indicator for work .................... 13 

6. Operating procedure......................................................... 14 

6.1. Measuring the mechanical stress of metal of steel parts 

and metal structures .......................................................... 14 

6.2. Measuring the tension of steel studs and bolts ............ 15 

6.3. Charging or replacing the battery................................ 16 

7. Maintenance service......................................................... 19 

8. Troubleshooting ............................................................... 20 

9. Transportation and storage rules....................................... 22 

Appendix. Mechanical and magnetic parameters of frequently 

arrived Russian grades of structural steels ............................ 23 

 

 

This operating manual allows you to acquaint you with 

the  device  and  operation  of  the  IN-01m  metal  mechanical 
stress  indicator (hereinafter  referred  to  as the  stress  indicator) 
and sets the rules for its operation, transportation and storage, 
compliance  with  which  ensures  that  it  is  maintained  in 
constant readiness for operation. 
 
 

1. PURPOSE 

 

1.1.    The  IN-01m  metal  mechanical  stress  indicator  is 

intended  for  indirect  estimation  of  the  value  of  the  metal 
mechanical  stress  of  a  local  section  of  a  steel  product  or  a 
metal  structure  made  of  structural  steel  by  the  value  of  the 
field strength of the residual magnetization of the metal. 

1.2.  The  stress  indicator  can  also  be  used  to  measure 

the  tightening  force  of  steel  studs  and  bolts  of  housing 
equipment. 

1.3. The stress indicator can be used in field, workshop 

and laboratory conditions. 

1.4.  Sources  of  electromagnetic  industrial  interference 

must  be  removed  from  the  stress  indicator  at  a  distance  of  at 
least 3 m. 

1.5.  Degree  of  protection  against  penetration  of  solids 

and  water  for  the  stress  indicator  IP41  according  to  GOST 
14254-80. 

1.6. Display of measurement results are digital  in units 

of  the  measured  value  of  the  field  strength  of  the  residual 
magnetization  of  the  metal  (mT),  or  the  mechanical  stress  of 
the metal (MPa). 

1.7. 

Indication 

of 

readouts 

is 

liquid 

crystal 

alphanumeric display with backlight for working in conditions 
the natural environment lighting is not enough. 

Summary of Contents for IN-01m

Page 1: ...Spector Sci Tech LLC IN 01m METAL STRESS INDICATOR USER MANUAL Russia Ufa 2020 2...

Page 2: ...1 PURPOSE 1 1 The IN 01m metal mechanical stress indicator is intended for indirect estimation of the value of the metal mechanical stress of a local section of a steel product or a metal structure ma...

Page 3: ...rent consumed from a freshly charged PP3 element mA when working without the display backlight 7 0 5 when working with the display backlight 16 0 5 2 6 Operating time of stress indicator when powered...

Page 4: ...effect sensor 2 via a scale amplifier combined with a 4 6 7 5 3 2 1 8 8 low pass filter 4 to the input of the signal level exceeding light indication unit 5 and the input of the analog to digital conv...

Page 5: ...receiving service information from the computer 3 3 4 4 Connector 12 is used to connect the charger s network adapter to the electronic unit for recharging the battery battery 3 4 The purpose of the...

Page 6: ...lays the current date and time of stress indicator on the screen 12 3 5 Magnetizing unit 3 5 1 The magnetizing unit of the stress indicator is used for magnetizing the local metal section of a steel p...

Page 7: ...rmation first appears then the value of the measured signal of the Sensor is set see section 3 4 2 of current document 14 5 1 5 In low light conditions to improve the visibility of the stress indicato...

Page 8: ...ation according to step 6 2 1 6 2 2 Install the stress indicator sensor in the middle area of the magnetized end surface of the steel stud or bolt 6 2 3 To measure the field strength of the residual m...

Page 9: ...me N Measurement N End of file 18 The data file represents sequentially recorded measurement results see table 6 1 6 4 3 Transfer of data recorded in the stress indicator s electronic memory to a pers...

Page 10: ...In the event of failure of the electronic components of the stress indicator they must be replaced 7 5 Current repairs are made during the operation of the stress indicator At the same time faults de...

Page 11: ...must ensure their stable position exclude the possibility of collision with other objects as well as against the walls of the vehicle 9 5 Conditions of transportation temperature C from minus 30 to p...

Page 12: ...11 0 11 5 12 14 5 650 400 5 6 8 0 8 5 10 11 Table A 2 r mT Steel grade B MPa Y MPa 0 Y B 3 350 210 0 62 1 69 2 0 09 2 470 325 1 06 1 38 2 43 2 79 2 98 17 1 520 350 1 38 3 12 4 12 Steel 20 420 230 1 3...

Reviews: