Sota Bio Tuner BT8 Instruction Manual Download Page 8

The Bio Tuner Model BT8 Manual

© 2012-2018 SOTA Instruments Inc. 

Page 8 of 9

|

 www.sota.com

ELECTRICAL SPECIFICATIONS

POWER REQUIREMENTS

Use a 9–Volt Alkaline battery for proper operation.

OPTIONAL POWER INPUT

The Bio Tuner can also be operated using rechargeable batteries. When purchasing 

rechargeable batteries, it is suggested to take the unit with you to ensure the batteries 

will fit the enclosure. 

OUTPUT SPECIFICATIONS

Maximum <1.5mA Peak ~40 Volts Peak-Peak into typical 4k ohm load. Minimum 

<~10uA, 10 Volt Peak-Peak at lowest setting. Output impedance ~20k ohm.
1,000 Hz (+/- 1%) pulse-modulated by 111.11 Hz (+/- 1%). Waveforms are Quartz 

Crystal-Controlled for precise frequency output. There is also a broad range of 

harmonic frequencies delivered in each pulse as observed on a frequency domain 

spectrum analyzer.

BIO TUNER MODES

MODE 1-3 (550 uS pulse width)

Mode 1: 1,000 Hz modulated by 111.11 Hz, 

Output = Continuous
Mode 2: 1,000 Hz modulated by 111.11 Hz, 

Output = 1 Second ON, 1 Second OFF (0.5 Hz)
Mode 3: 1,000 Hz modulated by 111.11 Hz,   

Output = Amplitude Modulated  by 7.83 Hz Schumann (Earth) Sine Wave

MODE 4-6 (220 uS PULSE WIDTH) 

Mode 4: 1,000 Hz modulated by 111.11 Hz,   

Output = Continuous
Mode 5: 1,000 Hz modulated by 111.11 Hz,   

Output = 1 Second ON, 1 Second OFF (0.5 Hz)
Mode 6: 1,000 Hz modulated by 111.11 Hz,   

Output = Amplitude Modulated by 7.83 Hz Schumann (Earth) Sine Wave 

Summary of Contents for Bio Tuner BT8

Page 1: ...USE 3 CAUTIONS 3 SUMMARY OF LIGHTS 3 HOW TO USE THE BIO TUNER 4 How it Should Feel 6 Typical Use 6 Changing the Conductive Rubber Electrodes 7 OUTPUT OF SIX MODES 7 ELECTRICAL SPECIFICATIONS 8 Power...

Page 2: ...also be used together with any of the SOTA Products With proper care your Bio Tuner should provide you many years of trouble free use Please understand that results will vary BEFORE USING Please read...

Page 3: ...ical condition or seizures please consult your health care provider prior to using the Bio Tuner 2 In very rare instances individuals have reported the onset or worsening of tinnitus when using the un...

Page 4: ...vity 3 Slide the clip on the back of the Bio Tuner into the holding clip on the Arm Band as in image 2 4 Insert the cord into the jack on the unit as shown in image 3 The USB symbol on the plug should...

Page 5: ...her timed session Before resetting for a second timed cycle wet the Ear Clips again b To turn the unit OFF rotate the ON OFF switch counter clockwise until it clicks Do not force the ON OFF switch pas...

Page 6: ...e following a Wet the conductive rubber of the Ear Clips again b Apply conductive gel or aloe vera gel to the ear lobes c Clean the Conductive Rubber Electrodes d Replace the Conductive Rubber Electro...

Page 7: ...pecially helpful Stop Gradually Reduce the amount of time each day for the last week of use CHANGING THE CONDUCTIVE RUBBER ELECTRODES 1 Remove the Ear Clips from the Cord 2 Gently pull the black Condu...

Page 8: ...e modulated by 111 11 Hz 1 Waveforms are Quartz Crystal Controlled for precise frequency output There is also a broad range of harmonic frequencies delivered in each pulse as observed on a frequency d...

Page 9: ...can appear at electrodes TROUBLESHOOTING A Troubleshooting Guide for the Bio Tuner is available at www sota com manuals This Guide is a great tool to check if you are experiencing any issues with your...

Reviews: