TC6
®
Repair and Replacement
5-7
5-6. Door Closed Sensor
Replacement
NOTE
Before you replace the door closed sensor, be sure it is
functioning properly by following the procedure given in
paragraph 5-5. Also, check the
new sensor before
installing.
1. Turn the centrifuge power OFF and unplug the power cord.
2. Remove the two phillips screws located on each side of the
bezel. Carefully lift the bezel up to disengage it from the frame
(but do not disconnect wiring) and gently place the bezel so that
the p.c. boards are facing up.
3. Disconnect P/J 331 (refer to figure 5-2). Loosen the back-up nut.
Then, while holding the back-up nut, remove and discard the
old door closed sensor threading it out through the
bottom
of the
bracket.
4. Open chamber door. Remove the sensor cover plug from the
cabinet.
Figure 5-2. Door Closed Sensor Parts Location
CHAMBER DOOR
MAGNET
TOP OF CABINET
SENSOR
COVER PLUG
DOOR CLOSED
SENSOR
BACK-UP NUT
SENSOR
BRACKET
CONNECTOR
P/J331