Preventive maintenance
38
Ultrasonic Generator | DocNr.: 03-4655 R03.01 | 22.09.2015 | en-US
Faults
8.1.2 Faults occurring during operation
During operation, other faults which can be attributed to other influencing factors
may occur. Notification of faults detected by the ultrasonic generator is provided in
two ways:
An alarm tone sounds.
The
ERROR
indicator light goes on.
When the ultrasonic generator has notified you of a fault, the cause of the fault must
be eliminated.
However, the ultrasonic generator does not detect all faults. For this reason, note any
faults occurring during operation. If you discover a fault, the cause of the fault must
be eliminated. Continuing to treat a patient despite the presence of a fault may result
in life-threatening injury to the patient. If a fault directly endangers the patient, turn
off the ultrasonic generator immediately.
Fault
Possible cause
Remedy
Instrument not working.
Instrument cable is defective.
Replace the instrument cable.
Instrument not automati-
cally detected.
Instrument is defective.
Use a different instrument.
The
OK
indicator light
flashes during treatment.
The pressure applied to the tis-
sue by the sonotrode tip is too
high.
Cease treatment. Reapply the so-
notrode tip to the skin.
During treatment, an
alarm tone sounds and
the
ERROR
indicator light
goes on.
Instrument is overloaded.
Cease treatment. Use a different
instrument if necessary.
No activation tone
sounds.
Activation tone has been turned
off.
Turn the volume control to the
right.
Instrument cannot be
turned off.
Foot switch is jammed.
Turn off the ultrasonic generator.
Activate the foot switch to elimi-
nate the jam.
Replace the foot switch if neces-
sary.
Irrigation not working.
There is no more irrigation solu-
tion in the irrigation bottle.
Connect a new irrigation bottle.
Irrigation tubing spike is not
properly positioned in the irriga-
tion bottle.
Insert the irrigation tubing spike
properly in the irrigation bottle.
Check the drip chamber in the pro-
cess.
Irrigation tubing is not correctly
inserted in the irrigation pump.
Open the irrigation pump and re-
insert the irrigation tubing.
The roller clamp on the irriga-
tion tubing is closed.
Open the roller clamp.
Negative pressure is building up
in the irrigation bottle because
the vent on the drip chamber is
not open.
Open the vent on the drip cham-
ber.
Irrigation is improperly set.
Adjust irrigation.