Installation
DocNr.: 03-4655 R03.01 | 22.09.2015 | en-US | Ultrasonic Generator
21
Turning on the ultrasonic generator
5.3 Turning on the ultrasonic generator
Refer to the instructions for use of the instrument and the accessory before starting
the ultrasonic generator.
After being turned on, the ultrasonic generator runs through a self-test. During the
self-test, check that all indicator lights go on briefly.
Faulty initial start-up
Faulty initial start-up of the ultrasonic generator may result in life-threatening situa-
tions or cause significant material damage.
The ultrasonic generator may only be put into operation for the first time by per-
sonnel authorized to do so by the manufacturer.
Requirements
The ultrasonic generator is connected to the main power supply.
A foot switch is connected.
The ultrasonic generator has reached the permissible temperature for use (be-
tween 10 °C and 40 °C (50 °F and 104 °F)).
Procedure
1. Turn the volume control on the back of the ultrasonic generator all the way to the
right.
2. Turn on the ultrasonic generator using the power switch.
– The green power switch goes on.
– The ultrasonic generator runs through a self-test.
3. During the self-test, check that all indicator lights briefly go on and that a tone
can briefly be heard.
– All of the indicator lights briefly go on and the tone can briefly be heard.
– The self-test has been completed successfully.
– The ultrasonic generator is turned on.
out.fm Seite 21 Dienstag, 22. September 2015 3:24 15