Instructions for use Airon
Instructions for use Airon
31 of 36
2021-06-10
4
Repairs/maintenance/re-use
For safety reasons and to maintain product liability, an inspection by your specialist dealer is re-
quired at least once a year. This must be carried out and documented according to the checklist
on the following page.
4.9 Maintenance / inspection
4.6 Lifespan
The expected normal service life, depending on the intensity of use and the number of re-uses,
is 5 years. For this purpose, the product must be used for its intended purpose and intended
use, the specifications in the instructions for use must be followed and all maintenance and ser
-
vice intervals must be observed.
The product can be used beyond this period if it is in a safe condition. This usual, theoretical
service life is not a guaranteed service life and is subject to a case-by-case examination by the
specialist trade, as is the reusability.
Use beyond the specified service life leads to an increase in the residual risks and should only be
carried out after careful, qualified consideration by the operator.
The service life can also be shortened depending on the frequency of use, the environment in
which it is used and how it is maintained.
The usual service life does not refer to wearing parts such as textile parts, wheels and plastic
parts that are subject to material-specific aging and / or wear.
This specified service life does not constitute an additional warranty or guarantee.
The wheelchair may only be disposed of with the approval of the insurance provider. The
wheelchair must be disposed of in accordance with the applicable national legal regulations.
4.8 Disposal
4.7 Reuse
Before re-use, a complete inspection according to the checklist must be carried out by a quali-
fied specialist retailer as well as complete cleaning and disinfection. We recommend replacing
all upholstery and textile parts for use by a new user.