3-8 (E)
WLL-RX55
b
c
GND
a
WF OUTPUT
Waveform/Vector monitor
WF
REMOTE
WLL-RX55 Connector panel
CCA-5 CABLE
(max. 200 m)
Master serup unit
MSU-700A/750
3-3-2. GRBY Output Level Adjustment
n
Perform the adjustments in Sections 3-3-2 and 3-3-3
continuously.
Tools
.
Master setup unit:
MSU-700A, or MSU-750
.
Oscilloscope:
Tektronix TDS3054B or equivalent
.
Waveform/Vector monitor:
Tektronix 1750A or equivalent (for
NTSC)
Tektronix 1751A or equivalent (for
PAL)
.
Extension board:
EX-854
(Sony Part No.: A-8344-228-A)
Connection
Preparations
1.
Turn OFF the power of the unit, and connect the unit
and equipment as shown in the connection diagram.
2.
Extend the VA-220 board using the EX-854 board.
3.
Turn ON the power of the unit and peripheral equip-
ment, and warm them up for about ten minutes.
4.
Menu setting:
MENU:
Setup
ITEM:
WF SEQ
→
4
BARS
→
CB1
5.
MSU-700A/750 setting:
BARS button
→
ON (Lit)
G Output Level Adjustment
1.
G video level adjustment
Equipment:
Oscilloscope
Test Point:
TP202 (GND : E3)/VA-220 board
Adjustment Point:
1
RV201/VA-220 board
Specifications:
a = 500
±
10 mV p-p
2.
MSU-700A/750 setting:
WAVEFORM MONITOR button
→
G (Lit)
3.
WF level adjustment
Equipment:
Waveform monitor
Test Point:
WF OUTPUT connector
Adjustment Point:
1
RV401/VA-220 board
Specifications:
b = 100
±
1 IRE (for NTSC)
b = 700
±
10 mV (for PAL)
4.
Measure the DC level c of the G signal.
Summary of Contents for WLL-RX55
Page 1: ...WIRELESS CAMERA RECEIVER WLL RX55 MAINTENANCE MANUAL 1st Edition Serial No 100001 and Higher ...
Page 64: ......
Page 126: ......
Page 134: ......
Page 140: ......
Page 154: ......
Page 164: ......
Page 184: ......
Page 194: ......
Page 196: ......
Page 230: ...Printed in Japan Sony Corporation 2004 2 08 B P Company 2004 WLL RX55 SY J E 9 968 042 01 ...